期刊文献+

X射线荧光光谱法测定钴基溶液中的6种元素

Determination of six elements in cobalt based solution by X-ray fluorescence spectrometry
下载PDF
导出
摘要 利用低功率X射线荧光分析仪,探讨了影响钴基溶液中Ni、Cu、Fe、Co、Zn、Mn测量的因素,拟定了实验条件,试验数据表明:钴的RSD在0.24%~0.34%之间,镍的RSD在0.39%~4.42%之间,铜的RSD在1.67%~7.71%之间,铁的RSD在2.52%~3.90%之间,锌的RSD在4.00%~6.34%之间,锰的RSD在1.88%~4.03%之间,镍的加标回收率在96.0%~102.6%之间,铜的加标回收率在96.9%~104.8%之间,铁的加标回收率在96.4%~102.6%之间,锌的加标回收率在96.2%~101.6%之间,锰的加标回收率在97.0%~102.9%之间,完全满足分析要求。X荧光分析结果与原方法吻合较好,准确度高,分析速度快,实现了钴基溶液中Ni、Cu、Fe、Co、Zn、Mn的同时测定,满足分析要求。 using low-power X-ray fluorescence analyzer,the factors affecting the measurement of Ni,Cu,Fe,Co,Zn and Mn in cobalt based solution are discussed,and the experimental conditions are formulated.The experimental data show that the RSD of cobalt is between 0.24%~0.34%,that of nickel is between 0.39%~4.42%,that of copper is between 1.67%~7.71%,that of iron is between 2.52%~3.90%,that of zinc is between 4.00%~6.34%,that of manganese is between 1.88%~4.03%,and that of nickel is between 96.0%~102.6%,The recoveries of copper,iron,zinc and manganese are 96.9%~104.8%,96.4%~102.6%,96.2%~101.6%and 97.0%~102.9%,respectively,which fully meet the analysis requirements.The results of X-ray fluorescence analysis are in good agreement with the original method,with high accuracy and fast analysis speed.The simultaneous determination of Ni,Cu,Fe,Co,Zn and Mn in cobalt based solution is realized,which meets the analysis requirements.
作者 刘晓丽 赵勇 吴川眉 潘梅荣 王雨欣 LIU Xiao-li;ZHAO Yong;WU Chuan-mei;PAN Mei-rong;WANG Yu-xin(Testing Center Of Jinchuan Group Co.,Ltd.,Jinchang City,Gansu Province,731700)
出处 《世界有色金属》 2022年第10期149-152,共4页 World Nonferrous Metals
关键词 X射线荧光光谱仪 钴基 X-ray fluorescence spectrometer cobalt based nickel copper iron cobalt zinc manganese
  • 相关文献

参考文献1

二级参考文献9

共引文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部