期刊文献+

Aqueous ruthenium detection by microwave-assisted laser-induced breakdown spectroscopy

下载PDF
导出
摘要 Aqueous ruthenium was detected in real-time under ambient conditions using microwaveassisted laser-induced breakdown spectroscopy(MW-LIBS).A 10 mJ laser energy and 750 W microwave power were directed at an open liquid jet sample of ruthenium.It was observed that,for liquid flow,the coupling efficiency between the microwave and the laser-induced plasma was limited to 43%.The improvement in the ruthenium’s signal-to-noise ratio with MW-LIBS,with respect to LIBS,was 76-fold.Based on MW-LIBS,the limit of detection for aqueous ruthenium was determined to be 957±84 ppb.
出处 《Plasma Science and Technology》 SCIE EI CAS CSCD 2022年第8期32-39,共8页 等离子体科学和技术(英文版)