期刊文献+

离子色谱测定水汽中甲酸根和乙酸根不确定度评定 被引量:1

Evaluation of Uncertainty in the Determination of Formate and Acetate by Ion Chromatography
下载PDF
导出
摘要 利用离子色谱检测水汽中甲酸根和乙酸根含量,识别并评定了引起结果不确定度的分量。当水汽样品中甲酸根含量为13.6042μg/L时,其合成扩展不确定度为0.9348μg/L(k=2);当乙酸根含量为16.7426μg/L时,其合成扩展不确定度为0.8676μg/L(k=2)。在各不确定度分量中,以校准曲线拟合引起的不确定度对结果影响最为突出。 Use ion chromatography to detect formate and acetate of water vapor,identify and assess the uncertainty of the results of the component.When the content of formate in water sample was 13.6042μg/L,the expanded uncertainty of synthesis was 0.9348μg/L(k=2);when the acetate content was 16.7426μg/L,the synthetic expanded uncertainty was 0.8676μg/L(k=2).Among the uncertainty components,the uncertainty caused by calibration curve fitting is the most prominent one.
作者 张超 姬海宏 刘伟伟 李元斌 刘金 ZHANG Chao;JI Haihong;LIU Weiwei;LI Yuanbin;LIU Jin
出处 《计量与测试技术》 2022年第9期113-116,共4页 Metrology & Measurement Technique
关键词 离子色谱 甲酸根 乙酸根 不确定度 ion chromatography formate acetate uncertainty
  • 相关文献

参考文献6

二级参考文献40

共引文献154

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部