算法和射野设置对肺癌容积旋转调强计划的剂量学影响

Dosimetric effects of volumetric modulated arc therapy plans for lung cancer caused by different dose algorithms and radiation field settings
原文传递
导出
摘要 目的分析不同剂量计算算法和不同射野设置对肺癌容积旋转调强计划(VMAT)的剂量学差异,为临床计划设计提供参考。方法选择20例肺癌患者,分别设计4组VMAT计划:基于各向异性解析算法(AAA)的2野2弧(2F2A_AAA)、基于外照射光子剂量算法(AXB)射的2野2弧(2F2A_AXB)、基于蒙特卡罗算法(MC)的2野2弧(2F2A_MC)、基于MC算法的1野2弧(1F2A_MC)。分别对不同算法、不同射野设置的计划,在靶区覆盖、高量控制、剂量均匀性指数(HI)、适形性指数(CI),以及危及器官(OARs)受照剂量进行评估。结果3组不同算法的2F2A计划靶区结果表明,2F2A_MC在PGTV的D_(1%)和V_(95%)(受到95%处方剂量所包绕的靶区相对体积)上均优于2F2A_AAA(D_(1%):t=-2.44,P=0.03;V_(95%):z=-2.04,P=0.04)和2F2A_AXB(D_(1%):t=2.34,P=0.03;z=-3.21,P<0.01)。2F2A_AXB在PGTV的CI表现上优于2F2A_AAA(z=-3.66,P<0.01),与2F2A_MC相当。就危及器官而言,2F2A_AXB和2F2A_MC全肺的V_(5 Gy)上分别较2F2A_AAA减少了0.68%(z=-2.69,P=0.01)和3.05%(z=-3.52,P<0.01)。2F2A_AXB计划在全肺D_(mean)为1776.44 cGy,均优于2F2A_MC(t=2.67,P=0.02)和2F2A_AAA(t=8.62,P<0.01)。2F2A_AXB的Body_5mm在V20 Gy相较于2F2A_AAA和2F2A_MC分别减少了1.45%(z=-3.88,P<0.01)和2.01%(z=-3.66,P<0.01)。而不同射野设置的两组计划结果表明,1F2A_MC在PTV1的CI和PTV2的HI上均优于2F2A_MC(CI:t=2.61,P=0.02;HI:z=-2.20,P=0.03)。1F2A_MC在全肺D_(mean)相对于2F2A_MC增加了26.29 cGy(t=2.28,P=0.04)。结论在进行肺癌VMAT计划设计时,MC算法适用于靶区优先,AXB算法适用于危及器官优先;而仅有MC算法的情况下,靶区优先时推荐选择1F2A,危及器官优先时推荐选择2F2A。 Objective To analyze the dosimetric differences of volumetric modulated arc therapy(VMAT)plans for lung cancer caused by different dose calculation algorithms and radiation field settings and thus to provide a reference for designing clinical VMAT plans for lung cancer.Methods This study randomly selected 20 patients with lung cancer and divided them into four groups of VMAT plans,namely,a group adopting two fields and two arcs based on the AAA algorithm(2F2A_AAA),a group employing two fields and two arcs based on the AXB algorithm(2F2A_AXB),a group using two fields and two arcs based on the MC algorithm(2F2A_MC),and a group adopting one field and two arcs based on the MC algorithm(1F2A_MC).Then,this study evaluated the target coverage,high-dose control,dose homogeneity index(HI),conformity index(CI),and organs at risk(OARs)of the plans using different algorithms and radiation field settings.Results The planning target volume(PTV)results of two fields combined with two arcs(2F2A)of three groups using different algorithms are as follows.2F2A_MC achieved better results in both D_(1%)and V_(95%)(the relative volume of the target volume surrounded by 95%of the prescribed dose)of planning gross target volume(PGTV)than 2F2A_AAA(D_(1%):t=-2.44,P=0.03;V_(95%):z=-2.04,P=0.04)and 2F2A_AXB(D_(1%):t=2.34,P=0.03;z=-3.21,P<0.01).2F2A_AXB outperformed 2F2A_AAA(z=-3.66,P<0.01)and was comparable to 2F2A_MC in terms of the CI of PGTV.Regarding OARs,2F2A_AXB and 2F2A_MC decreased the V_(5 Gy) of the whole lung by 0.68%(z=-2.69,P=0.01)and 3.05%(z=-3.52,P<0.01),respectively compared to 2F2A_AAA.2F2A_AXB achieved a whole-lung D_(mean)of 1776.44 cGy,which was superior to that of 2F2A_MC(t=2.67,P=0.02)and 2F2A_AAA(t=8.62,P<0.01).Compared to 2F2A_AAA and 2F2A_MC,2F2A_AXB decreased the V20 Gy of Body_5 mm by 1.45%(z=-3.88,P<0.01)and 2.01%(z=-3.66,P<0.01),respectively.The results of the two groups with different field settings showed that 1F2A_MC was superior to 2F2A_MC in both the CI of PTV1 and the HI of PTV2(CI:t=2.61,P=0.02;HI:z=-2.20,P=0.03).Moreover,1F2A_MC increased the D_(mean)of the whole lung by 26.29 cGy compared to 2F2A_MC(t=2.28,P=0.04).Conclusions Regarding the design of VMAT plans for lung cancer,the MC algorithm is suitable for the target priority and the AXB algorithm is suitable for the OAR priority.When only the MC algorithm is available,it is recommended to choose 1F2A in the case of target priority and select 2F2A in the case of OAR priority.
作者 郑万佳 黎恩廷 黄思娟 朱韵婷 连锦兴 王明理 黄晓延 杨鑫 Zheng Wanjia;Li Enting;Huang Sijuan;Zhu Yunting;Lian Jinxing;Wang Mingli;Huang Xiaoyan;Yang Xin(Department of Oncology,Southern Theater Air Force Hospital of the People′s Liberation Army,Guangzhou 510050,China;Department of Biomedical Engineering,Guangzhou Xinhua College,Guangzhou 510520,China;Department of Radiotherapy,Sun Yat-sen University Cancer Center,State Key Laboratory of Oncology in South China,Collaborative Innovation Center for Cancer Medicine,Guangdong Key Laboratory of Nasopharyngeal Carcinoma Diagnosis and Therapy,Guangzhou 510060;Department of Neurology,Guangzhou Women and Children′s Medical Center,Guangzhou 510623,China;Department of Radiotherapy,The First Affiliated Hospital of Guangzhou University of Chinese Medicine,Guangzhou 510000,China)
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2022年第9期671-677,共7页 Chinese Journal of Radiological Medicine and Protection
基金 广东省基础与应用基础研究基金(2021A1515220140) 中山大学肿瘤防治中心“青年优创”计划(QNYCPY32) 肿瘤精准放疗登峰计划(2021-DF-009) 肿瘤精准放疗星火计划(2019-N-11-34)。
关键词 肺癌 容积旋转调强放疗 射野设置 剂量计算 Lung cancer Volume modulated arc therapy Irradiation field setting Dose calculation algorithms Dose difference analysis
  • 相关文献

参考文献3

二级参考文献14

  • 1范廷勇,李建彬,于金明,卢洁,刘娟,于会明,巩合义.MLC叶片宽度对IMRT计划的影响[J].肿瘤防治杂志,2005,12(8):611-613. 被引量:10
  • 2Ibbott GS, Followill DS, Molineu HA, et al. Challenges in credentialing institutions and participants in advanced technology multi-institutional clinical trials [ J ] Ant J Radiat Oncol Biol Phys, 2005,71 ( S1 ) : S71-S75.DOI : 10. 1016/j.ijrobp.2007.08.083.
  • 3Ezzell GA,Burmeister JW, Dogan N, et al. IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J] .Med Phys,2009,36( 11 ) :5359-5373.DOI : 10.1118/1.3238104.
  • 4Lafond C, Chajon E, Devillers A, et al. Impact of MLC leaf width on volumetric-modulated arc therapy planning for head and neck cancers [J].J Appl Clin Med Phys,2013,14(6) :4074.DOI: 10. 1120/jacmp.vl4i6. 4074.
  • 5van Kesteren Z, Janssen TM, Damen E, et al. The dosimetric impact of leaf interdigitation and leaf width on VMAT treatment planning in Pinnacle : comparing Pareto fronts [ J] .Phys Med Biol, 2012, 57 ( 10 ): 2943-2952. DOI: 10. 1088/0031-9155/57/ 10/2943.
  • 6Wang SC ,Gong YL,Xu QF,et al. Impacts of muhileaf collimators leaf width on intensity-modulated radiotherapy planning for nasopharyngeal carcinoma: analysis of two commercial elekta devices [ J ].Med Dosim, 2011,36 ( 2 ) : 153-159. DOI : 10.1016/j. meddos.2010.02.007.
  • 7邓瑾,郑荣辉,张秀萍,黄赖机,叶家才.前列腺癌调强放疗和三维适形放疗临床剂量学研究[J].肿瘤学杂志,2010,16(6):478-481. 被引量:7
  • 8王矫健,徐利明,全红,李长虎,阮长利.多叶准直器叶片到位偏差对调强放射治疗剂量的影响[J].中国医学物理学杂志,2013,30(3):4121-4124. 被引量:7
  • 9张富利,刘清智,王雅棣.基于配置不同类型MLC叶片加速器的宫颈癌术后调强放疗计划评估[J].中国医学物理学杂志,2013,30(5):4383-4386. 被引量:7
  • 10乔延伟.肿瘤精确放疗技术发展及应用现状[J].中国医疗设备,2014,29(7):73-76. 被引量:21