摘要 将ICP光谱仪可拆卸炬管优化改造为一体式炬管,制定方案及预算,实施改造后验证实际成效。验证过程中确定了适当检测条件,采用标准加入法进行定量,线性相关系数(R)为0.9992~0.9999,回收率为96.2%~106.3%,相对标准偏差(RSD)为0.92%~1.77%。证明改造后可以消除样品中基体体系效应、共存元素干扰、光谱干扰等对测定结果的影响,由此可以认为改造后检测精盐水各元素是准确的,改造是相对成功的。 The detachable torch tube of plasma spectrometer was optimized and transformed into an integrated torch tube.The plan and budget are formulated.The actual effect is verified after the transformation.Appropriate test conditions were determined during validation.The standard addition method was used for quantitative analysis.The results showed that the linear correlation coefficient(R)was 0.9992-0.9999,the recovery was 96.2%-106.3%,and the relative standard deviation(RSD)was 0.92%-1.77%.It can be proved that the influence of matrix system effect,coexisting element interference and spectral interference on the determination results can be eliminated after transformation.It can be concluded that the determination of elements in refined brine after transformation is accurate and the transformation is relatively successful.
作者 韦景文 WEI Jingwen(Guangxi Tiandong Jinsheng Chemical Co.,Ltd.,Baise 531500,China)
出处 《氯碱工业》 CAS 2022年第10期38-42,共5页 Chlor-Alkali Industry
关键词 ICP 改造 炬管 验证准确性 成本 ICP reform torch tube accuracy verification cost
  • 相关文献

参考文献4

二级参考文献20

  • 1陈新坤,1986年
  • 2忻新泉,计算机在化学中的应用,1986年
  • 3陈新坤,1985年
  • 4褚家成,分析化学,1984年,12卷,11期,961页
  • 5辛仁轩,电感藕合等离子体光源.原理装置和应用,1984年
  • 6匿名著者,1982年
  • 7李廷钧,发射光谱分析,1982年
  • 8夏道行,实变函数论与泛函分析.下,1982年
  • 9顾基发,计算数学与应用数学,1980年,1卷,28页
  • 10甘杰,许晶,罗岳平,曾欢欣,胡军.ICP-MS法同时测定地表水中18种金属元素[J].环境监测管理与技术,2010,22(5):36-38. 被引量:34

共引文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部