摘要 目的:比较常规乳腺BI-RADS分类与对比增强能谱X线摄影(CESM) BI-RADS再分类对乳腺病变及不同病理类型病变的诊断价值。方法:回顾性分析201例乳腺病变患者的病理及CESM图像。采用卡方检验分析强化程度及曲线对乳腺癌的诊断价值。采用卡方检验和Kappa值分析常规乳腺BI-RADS分类与CESM BI-RADS再分类对乳腺癌的诊断效能及一致性。比较乳腺BI-RADS分类与CESM BI-RADS再分类对乳腺各病理类型的诊断敏感度。结果:71.68%的乳腺癌呈明显强化,80.68%的良性病变中无强化或弱强化(P<0.001)。乳腺良、恶性病变中Ⅰ型分别占63.64%和6.19%,Ⅱ型为26.14%和17.70%,Ⅲ型为10.23%和76.11%(P<0.001)。CESM BI-RADS再分类对乳腺癌诊断效能优于乳腺BI-RADS分类(P<0.001),敏感度依次为94.69%、78.76%,特异度为89.77%、53.41%,符合率为92.54%、67.66%。浸润性乳腺导管癌、乳腺腺病、乳腺纤维腺瘤及乳腺导管内乳头状瘤亚组中,CESM BI-RADS再分类的诊断敏感度高于乳腺BI-RADS分类(97.80%vs.83.52%,P=0.001;81.82%vs.50.00%,P=0.016;93.33%vs.60.00%,P<0.001;85.71%vs.35.71%,P=0.016)。结论:CESM BI-RADS再分类综合乳腺病变形态学及功能学特征,明显提高钼靶对乳腺病变的诊断效能,对浸润性乳腺导管癌、乳腺腺病、乳腺纤维腺瘤及乳腺导管内乳头状瘤的诊断更有优势。 Objective:To compare the diagnostic efficiency of conventional mammography BI-RADS classification and contrast-enhanced spectral mammography(CESM) BI-RADS reclassification for breast lesions and pathological subtypes.Methods:The pathological and CESM images of 201 cases with breast lesions were retrospectively analyzed.Chi-square tests were used to evaluate the diagnostic accuracy of enhancement degree and pattern of breast lesions.Kappa index and chi-square test were used to analyze the consistency and diagnostic efficiency of conventional BI-RADS classification and CESM BI-RADS reclassification for breast lesions.The diagnostic sensitivity of various histological subtypes was compared between the two groups.Results:71.68% of breast malignant lesions presented strong enhancement,no or weak enhancement can be seen in 80.68% of the benign lesion(P<0.001).Among the benign and malignant breast neoplasms,there were 63.64% vs 6.19% in pattern Ⅰ,26.14% vs.17.70% in pattern Ⅱ,and 10.23% vs.76.11% in pattern Ⅲ(P<0.001,respectively).The diagnostic efficiency of CESM BI-RADS was significantly higher than those of conventional BI-RADS(P<0.001),the sensitivity was 94.69% and 78.76%,specificity was 89.77% and 53.41%,accuracy was 91.33% and 67.66%,respectively.The diagnostic sensitivity of CESM BI-RADS was higher than that of conventional BI-RADS in the subgroups of invasive ductal carcinoma,adenopathy,fibroadenoma,and intraductal papilloma(97.80% vs.83.52%,P=0.001;81.82% vs.50.00%,P=0.016;93.33% vs.60.00%,P<0.001;85.71% vs.35.71%,P=0.016).Conclusion:CESM BI-RADS can integrate morphological and functional characteristics of breast lesions,and effectively improve the diagnostic efficiency of mammography for invasive ductal carcinoma,adenopathy,fibroadenoma,and intraductal papilloma.
作者 华蓓 王勇 刘斋 任庆云 全冠民 陈军 HUA Bei;WANG Yong;LIU Zhai(Department of Radiology and Nuclear Medicine,the First Hospital of Hebei Medical University,Shijiazhuang 050051,China)
出处 《放射学实践》 CSCD 北大核心 2022年第11期1380-1385,共6页 Radiologic Practice
关键词 乳腺肿瘤 乳腺X线摄影 对比增强能谱乳腺X线摄影 乳腺影像报告和数据系统 Breast neoplasms Mammography Contrast enhancement spectral mammography Breast imaging reporting and data system
  • 相关文献

参考文献9

二级参考文献29

共引文献211

同被引文献32

引证文献3

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部