期刊文献+

血清骨代谢标志物实验室检测基于生物学变异的允许总误差在可比性验证中的应用

Application of Allowable Total Error Based on Biological Variation in Comparability Verification of Serum Bone Metabolism Markers in Laboratory Detection
下载PDF
导出
摘要 目的通过分析血清骨代谢标志物实验室检测项目基于生物学变异的允许总误差(total error,TEa)质量指标与基于室间质量评价(external quality assessment,EQA)TEa质量指标在实验室可比性验证中的结果,评价基于生物学变异的TEa质量指标在血清骨代谢标志物室内、外可比性验证中的实用性。方法收集宝鸡市中心医院2021年11~12月血清骨代谢标志物试剂更换批次可比性验证数据,同时收集相同项目2021年第2次实验室间比对数据,计算相对偏差。项目包括骨钙素、总Ι型胶原氨基端延长肽(TP1NP)、β-胶原特殊序列(β-CTX)、甲状旁腺素(PTH)和25-羟维生素D3[D3(25-OH)]。通过王治国主编《临床检验生物学变异与参考区间》附录获取血清骨代谢标志物基于生物学变异的TEa质量指标,以国家卫生健康委临床检验中心2021年第2次EQA反馈报告,获取EQA的TEa质量指标,将可比性验证数据用此两种来源质量指标做符合要求率判断,以P<0.05为差异有统计学意义。结果以基于EQA或专家推荐的TEa为质量标准,血清骨代谢标志物试剂更换批号、室间比对的5个项目的符合要求率均为100%,可比性验证全部通过。以基于生物学变异TEa的最低质量标准判断,试剂更换批号、室间比对符合要求率80%~100%,两种来源判断标准结果比较符合要求率的差异无统计学意义(χ^(2)≥1.000,均P>0.05),可比性验证通过。以基于生物学变异TEa适当质量标准判断,则5个项目符合要求率为20%~100%,β-CTX,骨钙素试剂更换批号比对符合要求率分别为20%和40%,差异有统计学意义(χ^(2)=0.048,0.246,均P<0.05),可比性验证不能通过;此时室间比对的符合要求率60%-100%,差异无统计学意义(χ^(2)=0.444~1.000,均P>0.05),但符合率60%的β-CTX可比性验证不能通过。结论血清骨代谢标志物实验室检测基于生物学变异的TEa质量指标可以通过生物学变异数据库获得[D3(25-OH)除外],有3级性能指标,更科学、更严谨,建议实验室首选应用。 Objective To verify the comparability between the total error(TEa)quality index based on biological variation and the TEa quality index based on external quality assessment(EQA)in the laboratory testing items of serum bone metabolism markers.The results in this study were used to evaluate the utility of biological variation-based TEa quality indicators in the validation of serum bone metabolism markers in vitro and in vivo comparability.Methods The comparability verification data of replacement batches of serum bone metabolism marker reagents in Baoji Central Hospital from November to December 2021 were collected,and the second inter-laboratory comparison data of the same project in 2021 was collected to calculate the relative deviation.Items include osteocalcin,total collagen type 1 amino-terminal extension peptide(TP1NP),β-collagen specific sequence(β-CTX),parathyroid hormone(PTH)and 25-hydroxyvitamin D3[D3(25-OH)].Obtain the TEa quality index of serum bone metabolism markers based on biological variation through the appendix of Clinical Testing Biological Variation and Reference Interval edited by WANG Zhiguo,and obtain the TEa quality index of EQA based on the second EQA feedback report of the National Center for Clinical Laboratory in 2021.The comparability verification data were judged by these two source quality indicators as the compliance rate,and P<0.05 was considered to be statistically significant.Results With the quality standard based on EQA or TEa recommended by experts,the compliance rates of the five items of serum bone metabolism marker reagent replacement batch number and inter-laboratory comparison were all 100%,and the comparisons were all passed.Judging by the minimum quality standard based on biological variation TEa,the compliance rate of reagent replacement batch number and inter-laboratory comparison was 80%~100%,and the difference in compliance rate between the two source judgment standards was not statistically significant(χ^(2)≥1.000,all P>0.05),the comparison was passed.Judging by the appropriate quality standard based on biological variation TEa,the compliance rate of 5 items was 20%~100%,and the compliance rates ofβ-CTX and osteocalcin reagent replacement batch numbers were 20%and 40%,respectively,with the differences were statistically significant(χ^(2)=0.048,0.246,all P<0.05),the comparison failed.The compliance rate of the inter-laboratory comparison was 60%~100%,and the differences were not statistically significant(χ^(2)=0.444~1.000,all P>0.05),but theβ-CTX alignment with a coincidence rate of 60%could not pass.Conclusion The TEa quality indicators based on biological variation in the laboratory detection of serum bone metabolism markers could be obtained through the biological variation database[except D3(25-OH)].There are three-level performance indicators,which are more scientific and more rigorous.Recommended laboratory preferred application.
作者 高自颖 何玲鸽 甄拴平 GAO Zi-ying;HE Ling-ge;ZHEN Shuan-ping(Department of Clinical Laboratory,Baoji Central Hospital,Shaanxi Baoji 721008,China;Department of Clinical Laboratory,Baoji Hospital of Traditional Chinese Medicine,Shaanxi Baoji 721001,China)
出处 《现代检验医学杂志》 CAS 2022年第6期192-197,共6页 Journal of Modern Laboratory Medicine
关键词 骨代谢标志物 允许总误差 可比性验证 bone metabolic markers total allowable error comparability verification
  • 相关文献

参考文献1

二级参考文献6

  • 1Nevalainen D, Berte L, Kraft C, et al. Evaluating laboratory perform- ance on quality indicators with the six sigma scale [ J 1. Arch Pathol Lab Med, 2000, 124(4) : 516-519.
  • 2WS/T403-2012.临床生物化学检验指标分析质量指标[s].中华人民共和国卫生部,2012.
  • 3GB/T20470-2006.临床实验室室间质量评价要求[s].中华人民共和国国家质量监督检验检疫总局,中国国家标准化管理委员会,2006.
  • 4Schoenmakers CH, Naus A J, Vermeer H J, et al. Practical applica- tion of Sigma Metrics QC procedures in clinical chemistry [ J ]. Clin Chem Lab Med, 2011, 49 ( 11 ) : 1837-1843.
  • 5Boonen K J, Curvers J, Timmerman AA, et al. Trueness in the meas- urement of haemoglobin: consensus or reference method? [ J]. Clin Chem Lab Med, 2011, 50(3) : 511-514.
  • 6康凤凤,王薇,何法霖,王治国.已通过ISO 15189认可的临床实验室常规生化指标的性能评价[J].临床检验杂志,2013,31(1):57-59. 被引量:15

共引文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部