模拟手和电阻电容元件在中频治疗仪传导发射试验中的应用研究 被引量:1

Analysis on the application of artificial hand and RC components in conducted emission test of medium frequency electrotherapy stimulator
下载PDF
导出
摘要 根据医用电气设备电磁兼容测试标准要求,中频治疗仪的患者耦合部分在进行传导发射试验时需接入模拟手和电阻电容(RC)元件。然而,现有测试标准对模拟手的连接方式阐述不够详尽。通过对工作原理不同的两种中频治疗仪进行传导发射测试,对比测试结果并分析,从而给出此类设备进行该项试验时模拟手和RC元件连接方式。 According to the requirements of the electromagnetic compatibility testing standard of medical electrical equipment,artificial hand and RC element should be connected to the patient coupling part of the medium frequency electrotherapy stimulator during the conducted emission test.However,the existing test standards do not describe the connection modes of the artificial hand in detail.By comparing and analyzing the results of conducted emission tests on two types of medium frequency electrotherapy stimulator with different working principles,the idea and suggestion of the connecting way of artificial hand and RC element in the test of this kind of equipment were proposed.
作者 王偲宇 孟志平 WANG Si-yu;MENG Zhi-ping(Electromagnetic Compatibility Office,Beijing Institute of Medical Testing,Beijing 101111,China;不详)
出处 《中国医学装备》 2022年第11期226-231,共6页 China Medical Equipment
关键词 模拟手 电阻电容(RC)元件 中频治疗仪 电磁兼容性 传导发射 Artificial hand Resistance-capacitance(RC)element Medium frequency electrotherapy stimulator Electromagnetic compatibility Conducted emission