摘要
采用盐酸硝酸混合酸分析样品,优化仪器测定参数,选取最佳工作条件,建立电感耦合等离子体原子发射光谱法测定粗二氧化碲中Cu、Ag、Bi、Se、Pb、Sb和As的分析方法,其测定分别为0.013%~4.69%、0.011%~0.45%、0.012%~5.35%、0.010%~4.78%、0.011%~4.89%、0.012%~5.18%、0.011%~5.11%。各元素检出限为0.0027mg/L、0.0045mg/L、0.0131mg/L、0.0528mg/L、0.0097mg/L、0.0217mg/L、0.0329mg/L。加标回收率为95.8%~109.4%。方法简便、准确、可靠,适用于粗二氧化碲中Cu、Ag、Bi、Se、Pb、Sb和As同时测定。
A method for the determination of copper,silver,bismuth,selenium,lead,stibium and arsenic in crude tellurium dioxide samples by inductively coupled plasma emission spectrometry(ICP-AES)was established.The samples were digested with mixtures of hydrochloric acid,nitric acid.The result showed that the measurement ranges for Cu,Ag,Bi,Se,Pb,Sb and As were 0.013%~4.69%,0.011%~0.45%,0.012%~5.35%,0.010%~4.78%,0.011%~4.89%,0.012%~5.18%and 0.011%~5.11%respectively.The detection limit of the above-mentioned seven elements were 0.0027 mg/L,0.0045 mg/L,0.0131 mg/L,0.0528mg/L,0.0097mg/L,0.0217mg/L and 0.0329mg/L.And the recoveries were 95.8%~109.4%.This method sample,precise and reliable,so that it can be used in the determination of Cu,Ag,Bi,Se,Pb,Sb and As in crude tellurium dioxide samples.
作者
刘春华
LIU Chun-hua(Fujian Zijing Mining and Metallurgy Testing Technology Co.,Ltd.,Longyan 364200,China)
出处
《世界有色金属》
2022年第22期143-145,共3页
World Nonferrous Metals