摘要
Combing the time corelated single photon counting(TCSPC)with fuorescence lifetime imaging microscopy(FLIM)provides promising opportunities in revealing important information on the microenvironment of cells and tissues,but the applications are thus far mainly limited by the accuracy and precision of the TCSPC-FLIM technique.Here we present a comprehensive in-vestigation on the perforance of two data analysis methods,the first moment(M_(1))method and the conventional least squares(Fitting)method,in quantifying fuorescence lifetime.We found that the Mp method is more superior than the Fitting method when the lifetime is short(70-400ps)or the signal intensity is weak(<10^(3) photons).