A超與IOL MASTER測量人工晶體度數的準確性比較

To Compare the Accuracy of A-scan and IOL Master in Intraocular Lens Power Measurement
下载PDF
导出
摘要 目的比較和分析白内障手術前A超與IOL MASTER測量人工晶體度數的準確性。方法選取本院過去半年病例,200例年齡相關性白内障患者,術前分別探用A超和IOL MASTER設備測量人工晶體度數,作為預測值。所有患者均接受白内障超聲乳化+人工晶體植入術,術中及術後1個月無任何併發症。術後1月檢測手術眼屈光度,並與術前A超或IOL MASTER測量的預測值做配對比較,分析術前預測值與術後1個月實際屈光度的差異。結果術前A超和IOL MASTER的預測值與手術後1個月實際屈光度的平均誤差分別為:0.40±0.37DS和0.41±0.35DS,統計學分析,兩組差異無顯著性(P>0.05)。結論A超和IOL MASTER對年齡相關性白内障患者術前預測值與術後1個月實際屈光度有良好的一致性,兩種檢測方法對患者術後屈光度誤差無明顯差異。 Objective To compare the accuracy of A-scan and IOL Master in intraocular lens power measurement.Methods Two hundred patients with age-related cataract were included in the study.Intraocular lens power was measured by A-scan and IOL-Master before surgery and calculated according to the SRK-T formula.e refractive outcome was measured by auto-refractometer one month after surgery.Resultse mean absolute refractive error(MAE)in A-scan was(0.40±0.37)DS,while that of IOL-Master was(0.41±0.35)DS.ere was no signicant dierence between the two methods(P>0.05).Conclusionere is no signicant dierence between A-scan and IOL Master in intraocular lens power measurement.
作者 方守諾 鄧麗賢 袁淑敏 熊倩瑜 陳濱暉 賴一凡 FONG Sao Nok;TANG Lai In;UN Sok Man;HONG Sin U;CHAN Pan Fai;LAI Iat Fan(Department of Ophthalmology,Kiang Wu Hospital,Macao,China)
出处 《镜湖医学》 2018年第1期18-19,共2页 MEDICAL JOURNAL OF KIANG WU
关键词 年齡相關性白内障 人工晶狀體 A超 IOL Master Age-related cataract Intraocular Lens A-scan IOL Master
  • 相关文献

参考文献1

二级参考文献7

  • 1李筱荣.白内障与人工晶状体.第1版.北京:人民卫生出版社2011:122.
  • 2Federman JL, Schubert HD. Complications associated with the use ofsilicone oil in 150 eyes after retina - vitreous surgery. Ophthalmologe1988;95(7):870-876.
  • 3Assi A,Woodruff S,Gotzaridis E,et al. Combined phacoemulsificationand tranpupillary drainage of silicone oil : resuls and complications. Br JOphthalmol 2001 ;85(8) :942-945.
  • 4Rose LT, Moshegov CN. Comparison of the zeiss 10LMaster andapplanation A-Scan ultrasound : biometry for intraocular lens calculation.Clin Experiment Ophthalmol 2003 ;31 (2) :121-124.
  • 5Murray DC, Potamitis T, Good P, et al. Biometry of silicone oil -filledeye.Eye(Lond) 1999; 13(Pt 3a) :319-324.
  • 6Bencic G,Yatavuk Z,Marotti Mal. Comparison of A-scan and MRIfor the measurement of axial length in silicone oil - filled eyes. Br JOphthalmol 2009;93(4) :502-505.
  • 7Takei K, Sekine Y, Okamoto F, et al. Measurement of axial length ofeyes with incomplete filling of silicone oil in the vitreous cavity using x-ray computed tomography. Br J Ophthalmol 2002;86( 1) :47-50.

共引文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部