期刊文献+

聚合酶链反应分析仪校准结果不确定度的评定

Evaluation of Uncertainty of Calibration Results of Dissolution Tester
下载PDF
导出
摘要 聚合酶链反应分析仪(PCR仪)在疾控、机关病毒检测、医院、药厂等多个领域有着广泛的应用。PCR仪器的准确度与校准结果直接相关,有助于计量校准技术的研究,能够保证数据的准确可靠,保持正常的仪器工作状态,同时符合相关实验室认可的要求。因此对PCR仪温度示值误差不确定度的来源进行了分析,并对各不确定度分量进行计算,给出了相应的扩展不确定度。 Polymerase chain reaction analyzer(PCR) is widely used in disease control, organ virus detection, hospitals, pharmaceutical factories and other fields. The accuracy of the PCR instrument is directly related to the calibration results, which is conducive to the research of the metrological calibration technology, can ensure the accuracy and reliability of the data, maintain the normal working state of the instrument, and meet the requirements of the relevant laboratory accreditation. Therefore, the source of uncertainty of PCR instrument temperature indication error was analyzed, and the uncertainty components were calculated, and the corresponding expanded uncertainty was given.
作者 李巧艳 吕菁 马茹娟 LI Qiao-yan;LV Jing;MA Ru-juan(CCIC Northwest Metrology&Test Co.,Ltd.,Shaanxi Xi'an 710068,China)
出处 《广州化工》 CAS 2022年第24期109-111,共3页 GuangZhou Chemical Industry
关键词 计量 PCR仪 温度 不确定度 measurement PCR instrument temperature uncertainty