期刊文献+

人体代谢舱的新风混合均匀度性能表征

Characterization of fresh air mixing uniformity in human metabolic chamber
下载PDF
导出
摘要 设计了温湿度受控条件下的示踪气体法来验证人体代谢舱内新风混合均匀度及舱体密闭性能的试验方法,并通过实验验证了该试验方法的可行性,为评估人体代谢舱测量结果准确性提供了一定的数据支撑。 In this paper,a tracer gas method under temperature and humidity controlled conditions is designed to verify the mixing uniformity of fresh air in the human metabolic chamber and the sealing performance of the chamber body.The feasibility of this test method is verified through experiments,which provides certain data support for evaluating the accuracy of the measurement results of the human metabolic chamber.
作者 张爱亮 李金龙 崔体运 蔡泽仁 刘刚 Zhang Ailiang;Li Jinlong;Cui Tiyun;Cai Zeren;Liu Gang(Shanghai Institute of Measurement and Testing Technology)
出处 《上海计量测试》 2022年第6期8-12,共5页 Shanghai Measurement and Testing
关键词 计量学 间接测热法 人体代谢舱 示踪气体 均匀度 metrology indirect calorimetry the human metabolic chambers tracer gas uniformity