有创诱导降血氧法评估可孚指夹式脉搏血氧仪的准确度 被引量:1

Invasive Controlled Desaturation Study of Accuracy of the Cofoe Pulse Oximeter
下载PDF
导出
摘要 目的:根据ISO 80601-2-61:2017的要求,验证可孚指夹式脉搏血氧仪的准确度。方法:纳入12例志愿者,同时用受试设备和质控设备进行脉搏血氧饱和度(SpO_(2))监测,通过调节志愿者吸入空气和氮气的比例逐步降低其SpO_(2),分别在5个SpO_(2)区间内,记录受试设备数据并采集对应时间的动脉血样本测得SaO_(2)。比较受试设备SpO_(2)和动脉血SaO_(2)并进行统计。结果:以血气分析SaO_(2)为对照,受试设备SpO_(2)准确度为1.98%,符合SpO_(2)准确度在非运动条件下均方根差为≤3.0%的要求。其脉率检测结果与质控设备的心率对比,偏差为(1.19±1.33)次/min。试验过程无不良事件发生。结论:可孚指夹式脉搏血氧仪的准确度符合ISO 80601-2-61:2017以及中国YY 0784-2010的要求。 Objective:To validate the accuracy of Cofoe pulse oximeter according to ISO 80601-2-61:2017.Methods:A total of 12 volunteers were included the trail.The volunteers were connected with investigational device and quality control device,the quality control device pulse oxygen saturation(SpO_(2))of volunteers was gradually declined by regulating the inhalational air and nitrogen ratio in order to step down and maintain in the specific section.The data of investigational device was recorded and the corresponded blood sample were collected.The SpO_(2) data from investigational device and the arterial oxygen saturation(SaO_(2))data from arterial blood gas analysis were used for further analysis.Results:Contrast with the SaO_(2),the accuracy of SpO_(2) of the investigational device was 1.98%,which met the requirement that the accuracy of SpO_(2) being≤3.0%under non-exercise conditions.Contrast with the heart rate of quality control device,the pulse rate of experimental device deviation was(1.19±1.33)beats per minute.Conclusions:The SpO_(2) accuracy of Cofoe pulse oximeter meets the requirement according to ISO 80601-2-61:2017 and YY 0784-2010.
作者 王高坚 陈亦蛟 林姣姣 王方旭 周克 祝继洪 裘燕 周大春 WANG Gao-jian;CHEN Yi-jiao;LIN Jiao-jiao;WANG Fang-xu;ZHOU Ke;ZHU Ji-hong;QIU Yan;ZHOU Da-chun(The Department of Anesthesiology,Sir Run Run Shaw Hospital,Zhejiang University School of Medicine,Zhejiang Hangzhou 310016;The Department of Anesthesiology,the Affiliated of Medicine School,Ningbo Universuty,Zhejiang Ningbo 315020;Suzhou Care-Real Medical Technology Co.,Ltd,Jiangsu Suzhou 215000)
出处 《中国医疗器械信息》 2023年第5期125-127,共3页 China Medical Device Information
关键词 经皮血氧饱和度 动脉血氧饱和度 准确度 pulse oxygen saturation arterial oxygen saturation degree of accuracy
  • 相关文献

参考文献1

二级参考文献8

  • 1Nellcor technical stuff. A technology overview of the Nellcor Oximax pulse oximetry system[EB/OL]. (2009-10-01) [2014-04-20]. http:// www.ece.mcmaster.calfaculty/debruin/Engineering% 20Education% 204BD4/Oximetry%20Notes/OxiMaxTechNote.pdf.
  • 2Dugani S ,Hodzovic I ,Sindhakar S ,et al. Evaluation of a pulse oxi- meter sensor tester[J]. J Clin Monit Comput ,2011,25(3) : 163-170.
  • 3Murkin J M ,Adams S J,Novick R J,et al. Monitoring brain oxygen saturation during coronary bypass surgery : a randomized, prospective study[J]. Anesth Analg, 2007,104 ( 1 ) : 51-58.
  • 4福禄克医疗测试.Index2患者模拟器[EB/OL].(2011-04-22)[2014-04-15]. http ://zh.flukebiomedical.com/biomedical/userdbiomedical- test/simulators-&-eontrollerslindex-2xl-series,htm?PID=56330,.
  • 5科讯医疗.FLUKEProSim系列生命体征模拟仪赢得与时间的赛跑[EB/OL].(2012-08-31)[2014-04-10].http://instrument.ofweek.com/2012-08/ART-8320088-8110-28638303.html.
  • 6刘红,莫国民,张欣.基于STM32的数字血氧饱和度检测系统设计[J].中国医学物理学杂志,2012,29(4):3533-3536. 被引量:6
  • 7刘婷,黄明,丛茂柠.透射式脉搏血氧饱和度检测系统的设计与实现[J].中国医疗设备,2012,27(12):55-58. 被引量:7
  • 8游海峰,谢勤岚.基于FPGA的脉搏血氧饱和度检测系统设计[J].计算机与数字工程,2013,41(2):222-225. 被引量:2

共引文献2

同被引文献5

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部