摘要 AT555非接触式眼压计是临床上比较先进的眼压计,是利用压平一定面积的角膜所需气体压力来测量眼压值。其在临床应用中能够帮助医生快速了解患者疾病,对患者后续诊治有很大帮助。为保证医疗设备能够正常运行,需要加强对眼压计的日常故障维修。文章概述了医疗设备日常故障维修的价值与存在的问题,提出了医疗设备的日常故障维修策略(主要包括加大日常保养与检查力度,加大岗位培训力度,不断提高医疗设备维修的整体水平等),并总结了AT555非接触式眼压计光路通道故障、强亮直线故障等的日常故障维修案例。 AT555 non-contact tonometer is an advanced tonometer in clinic.It uses the gas pressure required to flatten a certain area of cornea to measure the intraocular pressure.In clinical application,it can help doctors quickly understand patients’diseases and is of great help to patients’follow-up diagnosis and treatment.However,we should strengthen the daily fault maintenance of tonometer to ensure the normal operation of medical equipment.Firstly,this paper summarizes the value and existing problems of daily fault maintenance of medical equipment,puts forward the daily fault maintenance strategy of medical equipment:increase daily maintenance and inspection efforts,increase job training efforts,and continuously improve the overall level of medical equipment maintenance,and summarizes the daily fault maintenance cases of AT555 non-contact tonometer:optical path failure,strong bright straight line failure,etc..
作者 杜永峰 DU Yong-feng(Ninghe District Hospital of Traditional Chinese Medicine of Tianjin,Tianjin 301500)
出处 《中国医疗器械信息》 2023年第7期170-172,共3页 China Medical Device Information
关键词 AT555非接触式眼压计 日常故障维修 案例分析 医疗设备 光路通道故障 强亮直线故障 AT555 non-contact tonometer daily fault maintenance case analysis medical equipment optical path failure strong bright straight line failure
  • 相关文献

参考文献10

二级参考文献75

共引文献69

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部