摘要
目的 比较Sirius与A超测量SMART、SMILE和FS-LASIK三种手术前后中央角膜厚度(CCT)的差异性、相关性和一致性。方法 分别用Sirius及A超测量125例(249只眼)患者术前及术后3个月的CCT值。结果 三组患者术前Sirius和A超CCT值比较,差异均具有统计学意义(均P<0.001)。术后除SMART组外,Sirius和A超CCT值比较,差异具有统计学意义(P=0.755,P=0.04,P=0.000)。手术方式对术后两种仪器测量差异的影响具有统计学意义(F=28.504,P<0.001),且FS-LASIK组的差异明显大于SMART组和SMILE组(均P<0.001)。手术前后,两种仪器的相关性和一致性较好。结论 Sirius和A超在测量SMART、SMILE和FS-LASIK三种手术前后中央角膜厚度时显示了高度的相关性及一致性,结合使用可提高手术前后评估角膜参数的准确性。
Objective To compare the difference,correlation,and consistency of central corneal thickness(CCT)measured by Sirius corneal topography and A-scan ultrasound pachymetry before and after SMART,SMILE,and FS-LASIK.Methods The CCT of 125 patients(249 eyes)was measured by Sirius and A-scan ultrasound pachymetry before and 3 months after surgery.Results There were statistically significant differences in Sirius and A-scan ultrasound CCT values between the three groups(all P<0.001).Except for the SMART group,Sirius and A-scan ultrasound showed statistically significant differences in CCT values after surgery(P=0.755,P=0.04,P=0.000).The surgical method had a statistically significant effect on the postoperative difference between the two instruments(F=28.504,P<0.001).The difference was significantly more significant in the FS-LASIK group than in the SMART and SMILE groups(both P<0.001).The two instruments had good correlation and consistency before and after the operation.Conclusion Sirius and A-scan ultrasound measurements of the central corneal thickness of SMART,SMILE,and FS-LASIK patients before and after surgery showed high correlation and consistency.The use of Sirius and A-scan ultrasound together can improve the accuracy of corneal parameters both before and after surgery.
作者
李依静
袁正
周春阳
Li Yijing;Yuan Zheng;Zhou Chunyang(Eye School of Chengdu University of TCM,Chengdu,Sichuan,610075;Yinhai Eye Hospital Affiliated to Chengdu University of TCM,Chengdu,Sichuan,610084)
出处
《中医眼耳鼻喉杂志》
2023年第2期61-64,73,共5页
Journal of Chinese Ophthalmology and Otorhinolaryngology
基金
四川省自然科学基金(编号:2022NSFSC0594)。