期刊文献+

一种抗原稳定剂的研究及应用

Research and application of an antigen stabilizer
下载PDF
导出
摘要 目的 本文研究了试剂盒抗原单组份不稳定性的影响因素;方法 通过抗原缓冲液离子强度、pH、不同试剂瓶材质和添加阻断剂等方法,测定不同条件的抗原溶液37℃加速后发光值的降幅,判断抗原不稳定的主要原因。结果 在抗原溶液中加入高分子聚合物泊洛沙姆12600、14600,抗原37℃热加速10 d后,稳定性得到明显的提升,发光值降幅由60%降低至8%。结论 高分子聚合物对抗原聚集有抑制作用,维持抗原结构的稳定。 【Objective】 To study the influencing factors of the instability of the single component of the antigen in the kit.【Methods】 Through the methods of antigen buffer ionic strength,pH,different reagent bottle materials and adding blocking agents,the antigen solutions under different conditions were determined at 37℃,the decrease of the luminescence value after acceleration was used to determine the main reason for the instability of the antigen.【Results】 After adding the high molecular polymer poloxamer 12600 and 14600 to the antigen solution,after the antigen was heated at 37°C for 10 days,the stability was significantly improved,and the decrease of luminescence value decreased from 60% to 8%.【Conclusion】 The macromolecular polymer has an inhibitory effect on antigen aggregation and maintains the stability of the antigen structure.
作者 王慧敏 杨阳 逯阵毫 闫淼 王玉堂 WANG Huimin;YANG Yang;LU Zhenhao;YAN Miao;WANG Yutang(Zhengzhou Autobio Biological Engineering Co.,Ltd.,Zhengzhou,Henan 450000,China)
出处 《中国医学工程》 2023年第4期16-20,共5页 China Medical Engineering
关键词 抗原 稳定性 泊洛沙姆 antigen stability poloxamer
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部