基于温度传感器的可穿戴呼吸检测装置研究

Research on wearable respiration detecting devices based on temperature sensor
下载PDF
导出
摘要 呼吸作为维持机体正常工作的基本生理活动,在临床上是评价身体状态的重要生理指标。本文利用温度传感器检测鼻腔周围呼吸气流的温度变化获取呼吸信号,设计了一种可穿戴呼吸检测装置。实验对比研究了鼻腔周围不同位置的检测结果,得到最佳检测位置处于口鼻之间区域的中上部。在正常呼吸下,该系统获得呼吸频率与手动计数的均方根误差(RMSE)为0.63,95%置信区间为[-1.54,0.87]。在快速呼吸下,检测极限达到30次/min,可以实现呼吸过速的检测。对于呼吸衰竭、呼吸暂停和深呼吸测量结果都表现出明显的差异特征,表明该装置具有用于临床呼吸监护的应用前景。 As a basic physiological activity to maintain body homeostasis,respiration is an important physiological index to evaluate physical status in clinical diagnosis.A temperature sensor is used to detect the temperature change of airflow around the nasal cavity to obtain breathing signals,a wearable breathing detection device is designed.The experiment is carried out to compare the detection results of different positions around the nasal cavity,showing that the middle-upper part between the nose and mouth is the best detection position for obtaining the respiratory signal.Under normal breathing,the root mean square error(RMSE)between the respiratory rate(RR)obtained by the system and manual counting is 0.63 with a 95%confidence interval of[-1.54,0.87].The detection limit reaches 30 times per minute for rapid breathing,which can realize the detection of tachypnea.For respiratory failure,apnea and deep breathing,the measurement results show obvious different characteristics,exhibiting that the device has great application prospects for clinical respiratory monitoring.
作者 范圣龙 鲁子鹏 黄祖博 郝英超 刘剑 FAN Shenglong;LU Zipeng;HUANG Zubo;HAO Yingchao;LIU Jian(School of Control Science and Engineering,Shandong University,Jinan 250061,China)
出处 《传感器与微系统》 CSCD 北大核心 2023年第6期66-69,共4页 Transducer and Microsystem Technologies
基金 国家重点研发计划资助项目(2020YFB1313900) 国家自然科学基金资助项目(61801256) 山东省自然科学基金资助项目(ZR2020MF024)。
关键词 呼吸信号 呼吸频率 传感器 无线传输 检测装置 respiratory signal respiratory rate(RR) sensor wireless transmission detecting devices
  • 相关文献

参考文献7

二级参考文献53

共引文献52

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部