2种测量角质层水分含量仪器的相关研究

Study on the consistency of two instruments for measuring the moisture content in stratum corneum
下载PDF
导出
摘要 比较Corneometer和MoistureMeterSC 2种仪器对角质层水分含量测量结果的一致性,为人体临床应用提供理论依据。在24名志愿者手臂内侧分别使用Corneometer和MoistureMeterSC进行角质层水分含量测量,通过Pearson相关系数对2种检测角质层水分含量的仪器进行相关性分析。通过Bland-Altman进行一致性分析,应用统计学方法对2种方法测量的结果进行差异性分析,同时用CV值评估设备的重复性。结果显示,2种仪器在涂抹区域及未涂抹区域测量角质层水分含量时具有良好的相关性,r在0.667~0.902,CV在6%以下,仪器重复性都较好;在未涂抹区域一致性较差,两者之间的差值距离0值较远,并且有显著性,在涂抹产品后一致性较好,Corneometer和MoistureMeterSC有较弱的一致性。Corneometer和MoistureMeterSC重复性都较好,并且具有良好的相关性,但是一致性较差,不能替代,使用时根据临床实际需求进行选择。 The consistency between Corneometer and MoistureMeterSC instruments in measurement of the moisture content of stratum corneum was analyzed,which could provide a theoretical basis for human clinical application.The moisture content of stratum corneum on the inner side of the arms of 24 volunteers was measured using Corneometer and MoistureMeterSC,and the correlation analysis of the two instruments was analyzed via Pearson’s correlation coefficient.Consistency analysis was performed by Bland-Altman.Difference analysis was carried out with statistical methods upon the results measured by the two instruments.The reproducibility of the devices was also assessed via CV values.The results show that,the two instruments have good correlation when measuring the moisture content of stratum corneum in the applied area and unapplied area,with r value of 0.667~0.902;CV is below 6%,indicative of good reproducibility;Corneometer and MoistureMeterSC have poor consistency for testing the unapplied area whereas good consistency for testing the applied area.Therefore,Corneometer and MoistureMeterSC have weak consistency.Both Corneometer and MoistureMeterSC are well reproducible.The two instruments have good correlation but poor consistency,so they cannot replace each other.When used,it should be chosen according to the actual clinical needs.
作者 徐韦韦 韩兴艳 XU Wei-wei;HAN Xing-yan(Fosun Cosmetics(Shanghai)Bio-Technology Co.,Ltd.,Shanghai 200030,China)
出处 《日用化学品科学》 CAS 2023年第6期14-17,44,共5页 Detergent & Cosmetics
关键词 角质层水分 相关性 一致性 moisture of stratum corneum correlation consistency
  • 相关文献

参考文献1

二级参考文献4

  • 1赵希刚,李坤成,李永忠.单次屏气电影法MRI评价左心室功能的初步研究[J].中国医学影像技术,2005,21(1):65-67. 被引量:12
  • 2Bland JM,Altman DG.Statistical methods for assessing agreement between two methods of clinical measurement.Lancet,1986,i:307-310.
  • 3Bland JM,Altman DG.Measuring agreement in method comparison studies.Statis tical Methodsin Medical Research,1999,8:135-160.
  • 4Najib Aziz,Michael RI,Sally SD,et al.Interpreting Method Comparison Studiesby Useofthe Bland-Altman Plot:Reflecting the Importance of Sample Sizeby Incorporating Confidence Limitsand Predefined Error Limitsinthe Graphic.Clinical Chemistry,2004,50 (11):2216-2222.

共引文献216

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部