期刊文献+

一种改进型放射外科准直器的设计和研究

Design and research of an improved radiosurgical collimator
下载PDF
导出
摘要 本文基于Hygeia公司GMX-1型钴(60Co)放射外科系统,采用蒙特卡罗模拟程序GEANT4设计了一款新型准直器并进行了实测分析。该准直器在锥形通孔内置锥形环孔聚焦结构。EBT3胶片测量结果表明,在放射源活度、源轴距、射野尺寸等保持不变的条件下,该改进型准直器的射野半影比原来减小17.6%,吸收剂量率是原来的2.12倍。该改进型准直器在中心点吸收剂量率、射野半影两个关键技术参数上具有显著优势。 In this paper,a new collimator based on Hygeia GMX-1 cobalt(60Co)radiosurgery system was designed and analyzed using the Monte Carlo simulation program GEANT4.The results of EBT3 film measurements showed that the field penumbra of the improved collimator was reduced by 17.6%and the absorbed dose rate was 2.12 times higher,while the source activity,source axis distance and field size remained unchanged.The collimator has significant advantages in the two key technical parameters of absorbed dose rate and penumbra.
作者 周云 陈蔚海 刘永彪 严南 孙向上 廖文涛 刘俊雅 蒲越虎 ZHOU Yun;CHEN Weihai;LIU Yongbiao;YAN Nan;SUN Xiangshang;LIAO Wentao;LIU Junya;PU Yuehu(Shanghai Institute of Applied Physics,Chinese Academy of Sciences,Shanghai 201800,China;University of Chinese Academy of Sciences,Beijing 100049,China;Medical Equipment Innovation Research Center,West China Hospital,Sichuan University,Chengdu 610041,China;Techtron information technology(Shanghai)Co.,Ltd.,Shanghai 200233,China;Nanjing Maicong Medical Technology Co.,Ltd.,Nanjing 210023,China;Department of Radiation Physics and Technology,Jiangsu Province Hospital,Nanjing 210029,China;Department of Radiation Physics and Technology,Hunan Cancer Hospital,Changsha 410013,China)
出处 《辐射研究与辐射工艺学报》 CAS CSCD 2023年第3期53-63,共11页 Journal of Radiation Research and Radiation Processing
基金 国家重点研发计划项目(2016YFCO105408)资助。
关键词 放射治疗 准直器 剂量率 射野半影 Radiotherapy Collimator Dose rate Penumbra
  • 相关文献

参考文献3

二级参考文献18

  • 1杨国君,张卓,陈思富,刘承俊,林郁正.渡越辐射在强流电子束诊断中的应用[J].高能物理与核物理,2005,29(3):305-311. 被引量:7
  • 2濑口忠男.聚合物的基本辐射效应:辐射向聚合物的能量转移[J].辐射研究与辐射工艺学报,2007,25(4):193-196. 被引量:6
  • 3Jung R, Ferioli G, Hutchins S. Single pass optical profile monitoring[C]. Mainz, Germany:Proceeding of DIPAC 2003, 2003:10-14.
  • 4Graves W S, Johnson E D, O'Shea P G. A high resolution electron beam profile monitor[C]. Vancouver:Proceeding of the 1997 Particle Accelerator Conference, 1997:1993-1995. DOI:10.1109/PAC.1997.751084.
  • 5Peter T, Tsumoru S. Measurement of small electron beam spots[J]. Annual Review of Nuclear and Particle Science, 1999, 49:125-162. DOI:10.1146/annurev.nucl.49.1.125.
  • 6Kube G, Behrens C, Gerth C, et al. Inorganic scintillators for particle beam profile diagnostics of high brilliant and high energetic electron beams[C]. New Orleans:Proceeding of the 2012 International Particle Accelerator Conference, 2012.
  • 7Blair G A, Kamps T, Lewin H, et al. R&D towards a laser based beam size monitor for the future linear collider[C]. Paris:Proceeding of EPAC 2002, 2002:1912-1914.
  • 8Russell S J, Carlsten B E. Measuring emittance using beam position monitors[C]. Washington, DC:Proceeding of the 1993 Particle Accelerator Conference, 1993:2537-2539. DOI:10.1109/PAC.1993.309381.
  • 9Floettamann K. DESY:ASTRA manual[OL]. 2000-9-18.
  • 10Chen Q, Chen Y, Chen M L, et al. A slit method to determine the focal spot size and shape of Tomo therapy system[J]. Medical Physics, 2011, 38(6):2841-2849. DOI:10.1118/1.3589133.