期刊文献+

核医学场所平面布局设计及效果分析

Plane layout design and effect analysis of nuclear medicine workplace
原文传递
导出
摘要 目的 设计核医学场所布局分区和通风方案并进行验证检测,为规范核医学场所的防护、保障核医学人员的辐射安全提供基础条件。方法 以2019—2022年北京市某医院设计并建成投入使用的核医学场所作为调查对象,通过现场勘察,查阅资料和综合分析,对该核医学场所的布局和通风进行设计,并在场所投入运行后进行放射性污染水平和通风效果验证检测,依据有关法规和技术标准进行分析评价。结果 确定了该核医学工作场所的布局分区和通风设计方案,验证检测结果表明,场所内废物桶表面放射性表面污染水平最高,为6.40 Bq/cm^(2),其次为便池和注射台,分别为0.70和0.50 Bq/cm^(2),场所内其他设施和物品表面以及现场工作人员体表均未发现明显的放射性污染;通风柜操作口截面风速为1.0 m/s。场所内表面污染水平和通风分别满足GB 18871—2002《电离辐射防护与辐射源安全基本标准》和GBZ 120—2020《核医学放射防护要求》的要求。结论 核医学场所布局分区和通风是防护设计的关键,验证检测是维护场所安全运行的有效手段,应将放射防护理念贯穿到每一个设计环节,从根本上为核医学场所运行的辐射安全提供可靠的保障。 Objective To design and verify the layout zone and ventilation scheme of nuclear medicine workplace,so as to provide basic condition for standardizing the protection of nuclear medicine workplace and safeguarding the radiation safety of nuclear medicine personnel.Methods A nuclear medicine workplace designed and built by a hospital in Beijing during 2019-2022 was taken as the object of investigation and study.The layout and ventilation of the nuclear medicine workplace were designed and evaluated according to relevant regulations and standards by means of site investigating,data retrieving and comprehensive analyzing.Furthermore,the radioactive contamination level and ventilation effect were verified and tested after the workplace is put into operation.Results The layout zone and ventilation design of nuclear medicine workplace were determined.The verification testing results showed that the highest level of radioactive contamination was 6.40 Bq/cm^(2)on the surface of waste buckets,followed by urinals and injection tables,which were 0.70 and 0.50 Bq/cm^(2),respectively.No obvious radioactive contamination was found on the surface of other facilities,articles and on-site staff in the workplace.The exhaust wind speed of the operation section of the fume hood was 1.0 m/s.The surface radioactive contamination level and ventilation in the nuclear medicine workplace met requirements of the Basic standards for ionizing radiation protection and radiation source safety(GB 18871-2002)and Radiological protection requirements for nuclear medicine(GBZ 120-2020),respectively.Conclusion The layout zone and ventilation of nuclear medicine workplace are the key to protection design,verification test is effective to maintain the safe operation of the workplace,and the concept of radiation protection should be integrated into every design link to provide reliable protection for the radiation safety of nuclear medicine workplace fundamentally.
作者 廖鑫 马永忠 LIAO Xin;MA Yong-zhong(Beijing Tsinghua Changgung Hospital/lnstrument Department of Clinical Medicine School of Tsinghua University,Bejing,102218,China;Institute for Radiation Hygiene Protection,Beijing Center for Disease Prevention and Control,Beijing,100013,China)
出处 《职业与健康》 CAS 2023年第11期1464-1468,共5页 Occupation and Health
基金 北京市高层次公共卫生技术人才建设项目(学科带头人-01-20)。
关键词 核医学场所 平面布局 通风 放射防护 Nuclear medicine workplace Plane layout Ventilation Radiation protection
  • 相关文献

参考文献10

二级参考文献58

  • 1陆克义,李险峰.甲状腺疾病患者^(131)I治疗后对他人的辐射危害评价[J].国外医学(放射医学核医学分册),2004,28(5):226-229. 被引量:9
  • 2赵兰才,张丹枫,张翼,等.放射防护使用手册[M].济南:济南出版社,2009:138-143.
  • 3中华人民共和国卫生部.GBZ 120-2006临床核医学放射卫生防护标准[S].北京:人民卫生出版社,2006.
  • 4Adedapo KS,Onimode YA,Ejeh JE,et al.Avoidable challenges of a nuclear medicine facility in a developing nation[J].Indian J Nucl Med,2013,28(4):195-199.
  • 5Guilabert N,Ricard M,Chamoulaud K,et al.Potential radiation hazard in nuclear medicine[J].Rev Prat,2015,65(1):83-84.
  • 6Adelstein SJ.Radiation risk in nuclear medicine[J].Semin Nucl Med,2014,44(3):187-192.
  • 7中国疾病预防控制中心辐射防护与核安全医学所.2013年放射诊疗建设项目职业病危害放射防护评价培训教材[M].北京:中国疾病预防控制中心辐射防护与核安全医学所,2013:196.
  • 8王雷,关立军.核医学科综合防护措施的调查与评价[J].中国地方病防治,2008,23(5):397-398. 被引量:4
  • 9巴建涛,付海鸿,吕京桥,邢海群,党永红.^131I治疗病房的辐射防护设计[J].中华放射医学与防护杂志,2009,29(6):614-617. 被引量:7
  • 10杨浩贤,梁绵英,谭光享,黄伟旭,郭玮珍.广东省核医学科工作场所放射卫生防护现状分析[J].现代生物医学进展,2010,10(15):2880-2883. 被引量:4

共引文献54

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部