摘要 建立超痕量六价铬分析仪的校准方法。基于超痕量六价铬分析仪的结构、工作原理,参考仪器出厂指标及JJG823—2014《离子色谱仪检定规程》,确定泵流量设定值误差和稳定性、基线噪声和漂移、最小检测浓度、定性和定量重复性作为校准项目。提出参考计量技术指标,泵流量设定值误差不超过±2%,泵流量稳定性不超过2%,基线噪声不超过0.025,30min基线漂移不超过0.25,最小检测质量浓度不大于0.05μg/L,定性重复性不大于1.5%,定量重复性不大于3%。通过校准实例对校准方法进行验证,证明建立的校准方法合理可行,可为超痕量六价铬分析仪的计量校准提供参考。 The calibration method of ultra-trace chromium(Ⅵ)analyzer was established.Based on the working principle and structure of the ultra-trace chromium(Ⅵ)analyzer,with reference to the instrument's factory index and the JJG 823—2014"Verification Regulation of Ion Chromatographs",the main calibration items were selected including pump flow setting value error and stability,baseline noise and drift,minimum detection concentration and measurement repeatability.The reference technical specifications were put forward,the pump flow setting value error should be within±2%,the flow stability should not be more than 2%,the baseline noise should not be more than 0.025,the baseline drift in 30 min should not be more than 0.25,the minimum detection mass concentration should not be more than 0.05μg/L,the qualitative repeatability should not be more than 1.5%and the quantitative repeatability should not be more than 3%.According to a calibration case,the calibration method was verified,and proved that the established calibration method was reasonable.This calibration method can be used as a reference for the calibration of the ultra-trace chromium(Ⅵ)analyzer.
作者 李萌 张娜 言思敏 LI Meng;ZHANG Na;YAN Simin(Shanghai Institute of Quality Inspection and Technical Research,Shanghai 201114)
出处 《化学分析计量》 CAS 2023年第6期85-89,共5页 Chemical Analysis And Meterage
关键词 超痕量 六价铬分析仪 校准方法 ultra-trace chromium(Ⅵ)analyzer calibration method
  • 相关文献

参考文献11

二级参考文献104

共引文献45

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部