期刊文献+

影响双干涉仪串联手持式FFOCT系统成像特性的主要因素分析

Main Factor Analysis for Imaging Performance of Hand-Held Full-Field Optical Coherence Tomography System with Dual Interferometers
原文传递
导出
摘要 报道一个改进的基于双干涉仪串联的手持式全场光学相干层析成像(FFOCT)系统。该系统中迈克耳孙干涉仪作补偿干涉仪,菲佐干涉仪作探测干涉仪,并在原有的系统基础上增设了一个平面反射镜以改变系统光路,便于样品的水平放置。分析了反射镜与其他系统主要元件的角度偏移对系统成像质量和系统补偿距离的影响。所获得的结果对手持式FFOCT系统的设计具有重要的指导意义。 In this study,we report an improved hand-held full-field optical coherence tomography(FFOCT)system based on a combination of double interferometers.The system uses a Fizeau interferometer for detection and a Michelson interferometer for compensation.The initial system was modified by the addition of a plane mirror to alter its optical path and to make it easier to arrange samples horizontally.The analysis was conducted on how the system’s imaging quality and compensation distance were affected by the angle deviation between the mirror and other key components of the system.The obtained results are useful for the design and optimization of the hand-held FFOCT system.
作者 李冉 高万荣 唐逸枫 Li Ran;Gao Wanrong;Tang Yifeng(School of Electronic and Optical Engineering,Nanjing University of Science&Technology,Nanjing 210094,Jiangsu,China)
出处 《激光与光电子学进展》 CSCD 北大核心 2023年第12期360-366,共7页 Laser & Optoelectronics Progress
基金 中央高校基本科研业务费专项资金(30920010003) 国家自然科学基金(61275198,60978069)。
关键词 全场光学相干层析术 成像系统 角度偏差 误差分析 full-field optical coherence tomography imaging system angle deviation error analysis
  • 相关文献

参考文献2

二级参考文献29

共引文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部