不同包装材料对外来器械灭菌升温过程的影响

Effect of different packaging materials on the heating process of loaned instrument sterilization
原文传递
导出
摘要 目的研究不同包装材料对外来器械灭菌过程中升温的影响,为精准灭菌提供参考。方法采用不同的包装材料对外来器械进行打包灭菌,每组分别测试7次,用温度压力检测仪监测温度、时间等参数。结果棉布、无纺布、硬质容器盒对外来器械灭菌过程的温度差异无统计学意义,灭菌平衡时间与灭菌维持时间存在部分布点[4号导向器导向道(钢质)、6号木制手柄内中心]差异有统计学意义(P<0.05);6号探头所在位置木制手柄内中心实测时间少于试验设定(300 s)。结论棉布相比硬质容器盒和无纺布,在外来器械的灭菌升温过程中存在延迟现象。 Objective To study the effect of different packaging materials on the heating process of loaned instrument sterilization,so as to provide reference for precise sterilization.Methods Different packaging materials were used to package and sterilize loaned instruments,and each group was tested for 7 times and parameters such as temperature and time were monitored with a temperature and pressure detector.Results There was no significant difference in the average sterilization temperatures of cotton cloth,non-woven fabric and hard container box.There were some distribution points in the sterilization balance time and sterilization maintenance time,such as guide channel of No.4 guide(steel)and inner center of No.6 wooden handle,and the difference was statistically significant(P<0.05);the actual measurement time of the inner center of the wooden handle where No.6 probe located was less than 300 s set in the experiment.Conclusion Compared with hard container box and non-woven fabric,cotton cloth has a delay phenomenon in the sterilization and heating process of loaned instruments.
作者 王俊锋 李漫春 詹朦 刘威 姚卓娅 耿军辉 丁丽娜 WANG Jun-feng;LI Man-chun;ZHAN Meng;LIU Wei;YAO Zhuo-ya;GENG Jun-hui;DING Li-na(Henan Provincial People's Hospital,Zhengzhou Henan 450003,China)
机构地区 河南省人民医院
出处 《中国消毒学杂志》 CAS 2023年第7期488-491,共4页 Chinese Journal of Disinfection
基金 河南省联合共建项目(SB201901089) 河南省医疗器械灭菌质控医学重点实验室开放课题项目(MJZK20220004)。
关键词 包装材料 外来器械 灭菌 升温 packaging materials loaned instruments sterilization heating
  • 相关文献

参考文献7

二级参考文献57

  • 1American National Standards Institute. ANSI/AAMI ST 79:2006 Comprehensive guide to steam sterilization and sterility assurance in health care {acilities [s]. American National Standards Institute,2006. 94.
  • 2Cofiell A,Hancock CO. An introduction to process validation, Purdue University/International Association for Central Service Material Management [M]. Self Study Series Lesson, 1998.41.
  • 3Schneider PM. Requirements for parametric release of steam sterilized products[M]. Central Sterilization, 2000.8.
  • 4Bryce EA, Robert FJ, Clernents B, et al. When the biological indicator is positive: investigating autoclave failures[J]. Infect Control Hosp Eiderniol, 1997,18(9) :654.
  • 5International Standards Organization. ISO 17665-1: Sterilization of healthcare products-Moist heat-Part 1, Requirements for the development, validation and routine control of a sterilization process for medical devices[s]. International Organization of Standardization, 2006.
  • 6International Standards Organization. ISO 17665-2. Sterilization of health care products-Moist heat-Part 2:Guidance on the application of ISO 17665-1[S]. International Organization of Standardization, 2009.
  • 7International Organization for Standardization. ENISO15882 Sterilization of health care products-Chemical indicators-Guid ance for the selection, use and interpretation of results[S] International Organization of Standardization, 2008.
  • 8International Organization for Standardization. ISO11139: Sterilization of health care products-Vocabulary[S]. International Organization of Standardization,2006.
  • 9中华人民共和国卫生部.WS 310.3-2009医院消毒供应中心第3部分:清洗消毒及灭菌效果监测标准[S],北京:中华人民共和国卫生部,2009.
  • 10WS/T367-201.医疗机构消毒技术规范[s].北京:中华人民共和国卫生部,2012.

共引文献71

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部