期刊文献+

Biplanar500E型G型臂X射线机故障维修

Troubleshooting of Biplanar500E G-arm X-ray machine
下载PDF
导出
摘要 阐述了Biplanar 500 E型G型臂X射线机的基本组成及工作原理,介绍了其在使用过程中出现的正位无法成像但侧位成像正常的原因及排除方法,为维修类似故障提供了参考。 The composition and working principle of Biplanar500E G-arm X-ray machine were described.The causes and elimination were introduced for the fault that imaging was abnormal at orthostatic position in case of normal lateral imaging.References were provided for treating similar faults.
作者 赵博文 魏世奇 ZHAO Bo-wen;WEI Shi-qi(Department of Medical Engineering,Hainan Hospital of Chinese PLA General Hospital,Sanya 572013,Hainan Province,China)
出处 《医疗卫生装备》 CAS 2023年第7期115-117,共3页 Chinese Medical Equipment Journal
关键词 Biplanar 500 E型G型臂X射线机 故障分析 故障维修 Biplanar500E G-arm X-ray machine fault analysis troubleshooting