期刊文献+

Sirius Scheimpflug-Placido versus. ultrasound pachymetry for central corneal thickness: meta-analysis

原文传递
导出
摘要 Background:To compare the difference in central corneal thickness(CCT)measurements in normal eyes between a rotating Scheimpflug camera combined with a Placido-disk corneal topographer(Sirius,CSO,Italy)and ultrasound pachymetry(USP).Methods:A systematic literature search was conducted for relevant studies published on PubMed,Medline,EMBASE,and the Cochrane Library and ClinicalTrials.gov from inception to August 1st,2019.Primary outcome measures were CCT measurements between Sirius and USP.A random effects model was used to pool CCT measurements.Results:A total of twelve studies involving 862 eyes were included in this meta-analysis.The meta-analysis found CCT measurements between Sirius and USP to be statistically significantly different(P<0.0001).The mean difference between Sirius and USP was-11.26μm with a 95%confidence interval(CI)(-16.92μm,-5.60μm).The heterogeneity Was IP=60%(P=0.004).Conclusion:CCT measurements with the Sirius Scheimpflug-Placido topographer were statistically significantly lower than USP.However,it may be argued that the mean difference of 11.26μm is not a clinically significant difference.
出处 《Eye and Vision》 SCIE CSCD 2023年第4期1-11,共11页 眼视光学杂志(英文)
基金 supported in part by the Foundation of Wenzhou City Science&Technology Bureau(Grant No.Y2020037) Medical and Health Science and Technology Program of Zhejiang Province(Grant No.2019KY111) Zhejiang Provincial&Ministry of Health Research Fund for Medical Sciences(Grant No.WKJ-ZJ-2134) Zhejiang Provincial High-level Talents Program(Grant No.2017-102).
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部