基于空间位移拉曼光谱的皮肤含水量无创检测

Noninvasive Detection of Skin Moisture Content Based on Spatially Offset Raman Spectroscopy
下载PDF
导出
摘要 基于空间位移拉曼光谱原理,搭建了一套可以实现微小空间位移的装置,确定了最优的空间位移距离为300μm,用于穿透皮肤表层,减少皮肤表面水份、油脂等污染物的干扰,最终采集到皮肤内层的水份拉曼信号。该装置采用拉曼波数在3390 cm^(−1)(水拉曼峰)和2935 cm^(−1)(蛋白峰)处的强度之比确定组织中的水-蛋白比例,采用便携式光谱仪使设备便于移动,探头和光谱仪以光纤相连使探头可灵活操作,设计满足临床使用要求。实验中检测了7个人、每人10个共70个不同位置的皮肤光谱,将光谱进行去背景算法处理,然后得到精确的拉曼峰强度。以电学法皮肤水份测试仪测试结果作为对比,以组内相关系数作为一致性分析指标,由单个测量组内相关系数为0.889可以看出,基于空间位移拉曼光谱法的人体活体皮肤含水量检测具有较高可行性。 In this study,the feasibility of detecting skin water content based on spatial displacement Raman spectroscopy and portable fiber optic spectrometer was explored.A device was constructed that can achieve an optimal spatial displacement distance of 300μm.It is selected to detect signals noninvasively under the surface of the skin and reduce the interference of pollutants such as water and oil on the surface of the skin.Instead of using the traditional spatial shift Raman spectroscopy device to achieve the relative shift of excitation point and collection point by mechanical offset or light blocking,a micro-shiftable light source was set up to adjust the spatial shift by using the conjugate relationship between the light source and the excitation point,and at the same time,a microscopic imaging device was introduced to control the adjustment,which finally realized the continuous adjustable spatial shift of less than 1000μm.The ratio of the intensity of Raman shift at 3390 cm^(−1)(water Raman peak)and 2935 cm^(−1)(protein Raman peak)is employed to characterize the skin moisture content,a portable spectrometer is employed to make the device moveable,the probe and the spectrometer are connected by optical fiber to make the probe flexible.Therefore,the design meets the clinical requirements.In total,70 spectra from different locations were collected(7 people,10 spectra each)in the experiment,and the spectra were processed by a de-backgrounding algorithm,and then the exact Raman peak intensity was obtained.The intra-group correlation coefficient was used as an index for consistency analysis.The intra-group correlation coefficient of 0.889 for a single measurement showed that the results of the two methods were in good agreement.In this study,four experiments with different spatial displacement amounts were set up,and the spatial displacement distances were 200μm,300μm,500μm and 1000μm.The skin spectra from the same person at the same location measured under different spatial displacement amounts can be seen to be significantly different only in signal intensity,while there is no significant and regular change in the spectral profile and details,and it is impossible to distinguish which spectrum is better The results of the quantitative analysis can only be judged by the results of the quantitative analysis.The experimental and analytical procedures described above were repeated at four spatial shifts in the study,and it was finally confirmed that the best results could be obtained at a spatial shift of 300μm.After proving the spatial sensitivity of the device by statistical methods and obtaining the optimal device parameters,we also hope to conclude a more intuitive method to quickly judge the signal superiority by spectral characteristics.As the spatial displacement distance increases,the spectral height decreases in turn,but the consistency shows a trend of first increasing,then decreasing and then increasing with the decreasing spectral height,there is no rule,so we cannot judge the signal superiority from the spectral height.We note that even for the same person at the same location,the relative intensity of the water signal(the ratio of the intensity of the Raman shift at 3390 cm^(−1)and 2935 cm^(−1))is not equal under the detection conditions of different spatial shift distances,the reason is that not all depths of the skin can correctly reflect the water content.Overall,if the relative intensity of the water signal is less than 0.05,then the detection results will only have low confidence,for example,when the spatial displacement distance is 500μm in the experiment,the relative intensity of the water signal is only 0.023,and the intra-group correlation coefficient is only 0.316.Under such conditions,it is necessary to adjust the optical path or check whether the skin of the subject has damage,scars and other problems;if the relative intensity of the water signal is close to 0.10,you can judge that the optical path configuration at this time is basically qualified,only need further fine tuning to determine the best parameters,the empirical method can quickly and effectively filter the weak signal,strong background spectrum,improve the efficiency of the study.It is proved that the experimental device can obtain a spectrum with good repeatability.The design of the fiber spectrometer and probe separation also meets the practical requirements,which is convenient for a large number of human body measurement experiments and has the potential to be transformed into medical devices.Indicating that the skin water content measured based on the spatial displacement Raman spectroscopy has high effectiveness and is expected to achieve convenient detection of the skin inner water content.
作者 袁榕穗 吴植强 聂荣志 钟文婷 张克勤 YUAN Rongsui;WU Zhiqiang;NIE Rongzhi;ZHONG Wenting;ZHANG Keqin(Guangxi Food and Drug Evaluation&Inspection Center,Nanning 530000,China;Guangxi Medical Device Testing Center,Nanning 530000,China;Department of Public Foundation Education,Hezhou University,Hezhou 542899,China;College of Textile and Clothing Engineering,Soochow University,Suzhou 215021,China)
出处 《光子学报》 EI CAS CSCD 北大核心 2023年第9期121-129,共9页 Acta Photonica Sinica
基金 国家自然科学基金(No.51873134)。
关键词 空间位移拉曼光谱 皮肤含水量 无创检测 一致性分析 组内相关系数 Spatially offset Raman spectroscopy Skin moisture content Noninvasive detection Correspondence analysis Interclass correlation coefficient
  • 相关文献

参考文献8

二级参考文献100

共引文献37

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部