摘要
目的:探讨医用直线加速器在实际应用中的质量状况,分析其使用中存在的问题,为质量管理提供科学依据。方法:根据GB/T19046-2013《医用电子加速器验收试验和周期检验规程》标准,对本院2台两种型号医用直线加速器,Varian6EX和VarianIX现场质量测试结果进行了分析。结果:本院医用直线加速器的初始合格率较高。经医疗机构质量控制人员和厂家工程师现场调试后,相关参数100%符合标准的要求。结论:必须加强医用直线加速器的质量控制措施,并定期进行性能检查,这确保了放射治疗的精确实施。
Objective:To explore the quality status of medical linear accelerators in practical applications,analyze the problems in their use,and provide scientific basis for quality management.Methods:According to the GB/T 19046-2013 Regulations for Acceptance Testing and Periodic Inspection of Medical Electron Accelerators,the on-site quality test results of two types of medical linear accelerators,Varian6EX and VarianIX,in our hospital were analyzed.Results:The initial qualification rate of our hospital’s medical linear accelerator is relatively high.After on-site debugging by quality control personnel from medical institutions and manufacturer engineers,the relevant parameters 100%meet the requirements of the standard.Conclusion:It is necessary to strengthen the quality control measures of medical linear accelerators and conduct regular performance checks to ensure the accurate implementation of radiation therapy.
作者
余育贤
吴继平
曾碧丽
张龙泉
YU Yu-xian;WU Ji-ping;ZENG Bi-li;ZHANG Long-quan(Zhangzhou Hospital Affiliated to Fujian Medical University,Fujian Zhangzhou 363000)
出处
《中国医疗器械信息》
2023年第19期81-83,143,共4页
China Medical Device Information
关键词
医疗器械
直线加速器
质量管理
medical devices
linear accelerator
quality control