摘要
为提高海洋环境放射性核素监测水平,开展了海水中^(137)Cs和^(90)Sr的实验室间测量比对活动。^(137)Cs测量结果与参考值的相对偏差为-2.99%~5.97%,测量结果的准确度、精密度和正确度均满足比对要求,比对评价结果均为“合格”;^(90)Sr测量结果与参考值的相对偏差为-41.58%~3.96%,其中12个实验室的比对评价结果为“合格”,1个实验室的比对评价结果为“不合格”。本次比对活动,各参比实验室的整体比对结果良好。
Inter-comparison results among analytical laboratories for the measurements of^(137)Cs and^(90)Sr in seawater in 2020 is described in this paper.The spiked seawater samples were distributed to 13 participating laboratories,and their analytical results were compared to the reference values.The relative deviation of^(137)Cs between measured values and reference value was-2.99%to 5.97%.The results had passed all criteria and were assigned“Accepted”status as a final score.In the case of^(90)Sr,one measured result did not pass the accuracy test and correctness test,so it was considered as“Not Accepted”.All other results were accepted.The analytical laboratories participating in this inter-comparison have good overall comparison results.
作者
林静
黄德坤
倪甲林
纪建达
钟强强
张金钊
于涛
LIN Jing;HUANG Dekun;NI Jialin;JI Jianda;ZHONG Qiangqiang;ZHANG Jinzhao;YU Tao(Laboratory of Marine Ecological Environment Early Warning and Monitoring,Third Institute of Oceanography,Ministry of Natural Resources,Fujian Xiamen 361005)
出处
《辐射防护》
CAS
CSCD
北大核心
2023年第5期485-489,共5页
Radiation Protection
基金
崂山实验室科技创新项目(LSKJ202202901)
自然资源部第三海洋研究所基本科研业务费专项资金资助项目(海三科2022005)。