摘要
A scintillator detector consisting of a LaBr_(3)(Ce)(0.5%)scintillator,a photomultiplier tube(PMT),and an oscilloscope were used to study the neutron sensitivities of the LaBr_(3)(Ce)scintillator at the China Spallation Neutron Source(CSNS)Back-n white neutron source in the double-bunch and single-bunch operation modes,respectively.Under the two operational modes,the relative neutron sensitivity curves of the LaBr_(3)(Ce)scintillator in the energy regions of 1–20 MeV and 0.5–20 MeV were obtained for the first time.In the energy range of 1–20 MeV,the two curves were nearly identical.However the relative neutron sensitivity uncertainties of the double-bunch experiment were higher than those of the single-bunch experiment.The above results indicated that the single-bunch experiment's neutron sensitivity curve has a lower minimum measurable energy than the double-bunch experiment.Above the minimum measurable energy of the double-bunch experiment,there is little difference between the measured relative neutron sensitivity curves of the single-bunch and double-bunch experiments of the LaBr_(3)(Ce)scintillator and those of other scintillators with a similar neutron response signal intensity.