使用Compass验证系统对脑瘤VMAT计划的三维剂量学进行定量评估

Quantitative Evaluation of Three-dimensional Dosimetry for Brain Tumor VMAT Program using Compass Verification System
下载PDF
导出
摘要 目的使用带有Dolphin探测器的Compass验证系统对脑瘤VMAT计划进行三维剂量验证,从而保证脑瘤放疗的剂量准确性以及临床更佳的剂量验证方案。方法选取20例在我院接受放射治疗的脑瘤患者,基于3%/2 mm评估标准,利用Compass独立计算和Dolphin实测重建结果比较靶区和危及器官的γ通过率。结果Compass独立计算结果显示,靶区γ通过率平均为94.92%、94.43%,危及器官γ通过率均在92%以上;Dolphin实测重建结果显示,靶区γ通过率平均为97.12%、96.67%,危及器官γ通过率则在93%以上。经证实二者均符合临床标准,但Dolphin实测重建结果优于Compass独立计算。结论在对脑瘤患者进行剂量验证时,尽管重建剂量结果优于独立计算结果,但是二者差异较小,基于三维剂量重建过程更加繁琐、耗时,临床上无法适用于所有患者,建议针对独立计算通过率较低的患者再进行实测,这样既能保证剂量准确性,又能减轻临床工作压力。 Objective To use the Compass verification system with Dolphin detector for three-dimensional dose verification of brain tumor VMAT plan,in order to ensure the accuracy of dose for brain tumor radiotherapy and a better clinical dose verification plan.Methods Selecting 20 brain tumor patients who received radiation therapy in our hospital,based on the 3%/2 mm evaluation standard,Compass independent calculation and Dolphin measured reconstruction results were used to compare the target area and organs at riskγpass rate.Results Compass independent calculation results display the averageγpassing rates of target area are 94.92%and 94.43%,the averageγpassing rates of organ at risk is above 92%;Dolphin measured reconstruction results show the averageγpassing rates of target area are 97.12%and 96.67%,the averageγpassing rates of organ at risk is above 93%.Research has found that both meet clinical standards,but Dolphin measured reconstruction results are better than Compass independent calculation.Conclusions When verifying the dose of brain tumor patients,although the reconstructed dose results are better than the independent calculation results,the difference between the two is relatively small.Based on the 3D dose reconstruction process,it is more cumbersome and time-consuming,and cannot be applied to all patients in clinical practice.This article suggests conducting further measurements for patients with lower independent calculation pass rates,which not only ensures dose accuracy but also reduces clinical work pressure.
作者 邱凌平 阙与清 李克 殷皓泽 张婷 李勇 QIU Lingping;QUE Yuqing;LI Ke;YIN Haoze;ZHANG Ting;LI Yong(Department of Oncology,the First Affiliated Hospital of Nanchang University,Nanchang 330006,China)
出处 《中国医药指南》 2023年第31期57-59,共3页 Guide of China Medicine
基金 新锐肿瘤支持治疗课题研究项目(cphcf-2022-035)(脑转移瘤立体定向放射治疗剂量验证方法的对比研究)。
关键词 脑肿瘤 剂量验证 Compass验证系统 Dolphin探测器 Brain tumor Dose verification Compass verification system Dolphin detector
  • 相关文献

参考文献5

二级参考文献49

  • 1Vanetti E,Clivio A,Nicolini G,et al.Volumetric modulated arc ra-diotherapy for carcinomas of the oro-pharynx,hypo-pharynx andlarynx:A treatment planning comparison with fixed field IMRT. Radiotherapy and Oncology .
  • 2Kjaer-Kristoffersen F,Ohlhues L,Medin J,et al.RapidArc volumet-ric modulated therapy planning for prostate cancer patients. Acta Oncologica . 2008
  • 3Ezzell GA,Galvin JM,LowD,et al.Guidance document on deliv-ery,treatment planning,and clinical implementation of IMRT:re-port of the IMRT Subcommittee of the AAPM Radiation Therapy Committee. Medical Physics . 2003
  • 4LIUHH,WANG X,DONG L,et al.Feasibility of sparinglung and other thoracic structures with intensity-modula-ted radiotherapy for non-small lung cancer. International Journal of Radiation Oncology Biology Physics . 2004
  • 5Imperato JP,Paleologos NA,Vick NA,et al.Effects of treatment on one-term surviors with malignant astrocytomas. Annals of Neurology . 1990
  • 6Esiashvili N,Koshy M,Landry J.Intensity-modulated radiation therapy. Current Problems in Cancer . 2004
  • 7VW Stieber,MP Mehta.Advances in radiation therapy for brain tumors. Neurologic Clinics . 2007
  • 8Palma D,Vollans E,James K, et al.Volumetric modulated arc therapy for delivery of prostate radiotherapy. Comparison with intensity-modulated radiotherapy and three-dimensional conformal radiotherapy. International Journal of Radiation Oncology Biology Physics . 2008
  • 9Cozzi L,Dinshaw KA,Shrivastava SK,et al.A treatment planning study comparing volumetric arc modulation with RapidArc and fixed field IMRT for cervix uteri radiotherapy. Radiotherapy and Oncology . 2008
  • 10Bertelsen A,Hansen CR,Johansen J,et al.Single Arc Volumetric Modulated Arc Therapy of head and neck cancer. Radiotherapy and Oncology . 2010

共引文献35

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部