原子喷泉中基于电光调制的激光移频与稳频

Laser Frequency Shift and Stabilization of Atomic Fountain Based on Electro-Optic Modulation
原文传递
导出
摘要 激光稳频是影响原子喷泉系统性能的关键技术。提出了一种应用于原子喷泉系统的激光稳频优化方案。该方案将激光冷却过程中采用的移频方法应用到稳频系统中,通过光纤电光调制的方法产生边带,并将频率锁定在边带的饱和吸收峰上。同时应用大光斑饱和吸收模块,改善饱和吸收信号和误差信号的信噪比。利用85Rb原子喷泉系统主光路的探测光,通过1.035 GHz的移频,实现从85Rb饱和吸收峰到87Rb饱和吸收峰的转移锁定,并在85Rb喷泉系统中观测到冷原子云信号。锁定后的信号可实现较长时间的稳定工作。该方案通过控制电光调制频率,还有望实现激光冷却实验范围内的任意频率锁定,具有重要的应用价值。 Objective In recent years,atomic frequency standards(AFSs)have been significantly improved and widely applied in measurement and transportation.The uncertainty of fountain AFSs has reached 10-16,and that of optical AFSs is much higher at 10-19,so AFSs are expected to replace Cs as a new definition of second.The development of AFSs is inseparable from the improvement of laser frequency stabilization.Narrow and ultra-narrow line-width lasers based on the frequency stabilization technology play significant roles in capturing and cooling atoms and ions,and trapping photons.Saturated absorption is widely employed in the frequency stabilization of AFSs because of its simple operation,high resolution,and the elimination of Doppler broadening.It is a technology commonly adopted in the microwave domain of atomic fountain clocks.The saturated absorption frequency stabilization of the atomic fountain clock is realized by locking a specific saturation absorption transition peak.The saturated absorption spectrum signal is modulated and demodulated to produce an error signal,and then a specific feedback signal is generated based on the error signal to control the laser output frequency.According to the output frequency,a series of modulation and frequency shifts are carried out to realize capturing,cooling,launching,and other functions which are required by the cold atom experiment.Generally,atoms have multiple saturated absorption peaks with only one involved in frequency locking,and unselected saturated absorption peaks cannot be utilized.Methods We propose an optimized method of laser frequency stabilization based on AFSs.This method applies the frequency control in laser cooling to the frequency stabilization system and realizes the transfer locking between different saturated absorption transition peaks.The sidebands are generated by fiber electro-optic modulation.Since the zero-order diffraction light cannot meet the frequency shift requirement,the fountain system generally takes+1(or-1)order diffraction light as the effective laser.We employ the-1-order diffraction light and adjust it to the maximum.Then the frequency on the saturated absorption peak of the sideband is locked to realize the laser frequency shift.A large aperture saturated absorption scheme is adopted to improve the signal-to-noise ratio(SNR)of the saturated absorption signal and its error signal after frequency shift.The feature of this scheme is to apply an adjustable aperture in the light path to improve the SNR of fringe by adjusting the size of the light spot.Finally,the signal differences in saturation absorption signals before and after modulation and optimization are compared.Results and Discussions The probing light of the main optical path of the 85Rb atomic fountain system employed for twolevel detection is adopted as the light source.The 1.035 GHz wide range frequency shift is realized by an optical fiber electro-optic modulator,which meets the requirements of the transition from F=3→F'=3⊗F'=4 of 85Rb to F=2→F'=3 of 87Rb.F=3→F'=3⊗F'=4 transition corresponds to the frequency locking position of 85Rb.The energy level transfer process is shown in Fig.2.We make the frequency-shifted laser pass through the saturated absorption optical path.Figure 4 shows the saturation absorption signal and the change of its SNR with the spot diameter.When the total input optical power remains unchanged and the spot diameter increases from 2 to 10 mm,the saturated absorption signal increases and its corresponding SNR increases by about 13 dB.Figures 4(c)and 4(d)indicate the relationship between error signal,spot diameter,and optical power.With the rising spot diameter and optical power(optical power density),the error signal and its slope increase with improved SNR.The changes in saturation absorption signal before and after modulation are shown in Fig.5.The saturated absorption signals before and after optimization are locked respectively.The signal before the optimization is too small to lock,while the optimized signal can be locked,andObjective In recent years,atomic frequency standards(AFSs)have been significantly improved and widely applied in measurement and transportation.The uncertainty of fountain AFSs has reached 10-16,and that of optical AFSs is much higher at 10-19,so AFSs are expected to replace Cs as a new definition of second.The development of AFSs is inseparable from the improvement of laser frequency stabilization.Narrow and ultra-narrow line-width lasers based on the frequency stabilization technology play significant roles in capturing and cooling atoms and ions,and trapping photons.Saturated absorption is widely employed in the frequency stabilization of AFSs because of its simple operation,high resolution,and the elimination of Doppler broadening.It is a technology commonly adopted in the microwave domain of atomic fountain clocks.The saturated absorption frequency stabilization of the atomic fountain clock is realized by locking a specific saturation absorption transition peak.The saturated absorption spectrum signal is modulated and demodulated to produce an error signal,and then a specific feedback signal is generated based on the error signal to control the laser output frequency.According to the output frequency,a series of modulation and frequency shifts are carried out to realize capturing,cooling,launching,and other functions which are required by the cold atom experiment.Generally,atoms have multiple saturated absorption peaks with only one involved in frequency locking,and unselected saturated absorption peaks cannot be utilized.Methods We propose an optimized method of laser frequency stabilization based on AFSs.This method applies the frequency control in laser cooling to the frequency stabilization system and realizes the transfer locking between different saturated absorption transition peaks.The sidebands are generated by fiber electro-optic modulation.Since the zero-order diffraction light cannot meet the frequency shift requirement,the fountain system generally takes+1(or-1)order diffraction light as the effective laser.We employ the-1-order diffraction light and adjust it to the maximum.Then the frequency on the saturated absorption peak of the sideband is locked to realize the laser frequency shift.A large aperture saturated absorption scheme is adopted to improve the signal-to-noise ratio(SNR)of the saturated absorption signal and its error signal after frequency shift.The feature of this scheme is to apply an adjustable aperture in the light path to improve the SNR of fringe by adjusting the size of the light spot.Finally,the signal differences in saturation absorption signals before and after modulation and optimization are compared.Results and Discussions The probing light of the main optical path of the 85Rb atomic fountain system employed for twolevel detection is adopted as the light source.The 1.035 GHz wide range frequency shift is realized by an optical fiber electro-optic modulator,which meets the requirements of the transition from F=3→F'=3⊗F'=4 of 85Rb to F=2→F'=3 of 87Rb.F=3→F'=3⊗F'=4 transition corresponds to the frequency locking position of 85Rb.The energy level transfer process is shown in Fig.2.We make the frequency-shifted laser pass through the saturated absorption optical path.Figure 4 shows the saturation absorption signal and the change of its SNR with the spot diameter.When the total input optical power remains unchanged and the spot diameter increases from 2 to 10 mm,the saturated absorption signal increases and its corresponding SNR increases by about 13 dB.Figures 4(c)and 4(d)indicate the relationship between error signal,spot diameter,and optical power.With the rising spot diameter and optical power(optical power density),the error signal and its slope increase with improved SNR.The changes in saturation absorption signal before and after modulation are shown in Fig.5.The saturated absorption signals before and after optimization are locked respectively.The signal before the optimization is too small to lock,while the optimized signal can be locked,and the atomic cloud signal is observed in the magneto-optical trap(MOT)area as shown in Fig.6.The locked saturated absorption signal can achieve stable operation for a long time and the error signal fluctuation is kept within 5×10-4 V,which meets the requirements of the atom fountain clock experiment.Conclusions In this study,we put forward an optimized method for laser frequency stabilization.The frequency of the incident light is shifted by electro-optical modulation to realize the transfer locking of 85Rb to 87Rb transition peaks through the probing light from the main optical path of the 85Rb fountain clock.The SNR of the transition spectral line is optimized by increasing the saturated absorption aperture,which has been improved by about 13 dB.After the frequency shift signal is locked,the atom cloud signal is observed on the atomic clock and can operate stably for a long time.By this method,we can achieve transfer locking of any frequency in the frequency range.It is also hopeful to optimize the main optical path by adding fiber electro-optical modulator(FEOM)in front of the main optical path and performing time control to realize some acousto-optic modulator(AOM)functions.As a result,the number of modulation devices is reduced and the optical path is simplified.Finally,the optical power transmission efficiency is improved to the possible realization of all fiber links.
作者 赵伟靖 李耀 董日昌 魏荣 Zhao Weijing;Li Yao;Dong Richang;Wei Rong(Key Laboratory for Quantum Optics,Shanghai Institute of Optics and Fine Mechanics,Chinese Academy of Sciences,Shanghai 201800,China;University of Chinese Academy of Sciences,Beijing 100049,China;Innovation Academy for Microsatellites of Chinese Academy of Sciences,Shanghai 200120,China)
出处 《光学学报》 EI CAS CSCD 北大核心 2023年第19期219-226,共8页 Acta Optica Sinica
基金 中国科学院战略性先导科技专项(B类)(XDB21030200) 国家重点研发计划(2021YFA0718303)。
关键词 原子喷泉 电光调制 移频 饱和吸收 稳定度 atom fountain electro-optic modulation frequency shift saturated absorption stability
  • 相关文献

参考文献6

二级参考文献42

  • 1栾广建,毛海岑,石晓辉.基于Rb87的调制转移光谱稳定激光器频率的方法[J].光学与光电技术,2020(2):83-86. 被引量:4
  • 2Bureau International Des Poids Et Meaures 2013 BIPM Annual Report on Time Activities (Pavillon de Breteuil).
  • 3Heavner T P, Donley E A, Levi F, Costanzo G, Parker T E, Shirley J H, Ashby N, Barlow S and Jefferts S R 2014 Metrologia 51 174.
  • 4Peil S, Hanssen J L, Swanson T B, Taylor J and Ekstrom C R 2014 Metrologia 51 263.
  • 5Jefferts S R, Shirley J, Parker T E, Heavner T P, Meekhof D M, Nelson C, Levi F, Costanzo G, Marchi A D, Drullinger R, Hollberg L, Lee W D and Walls F L 2002 Metrologia 39 321.
  • 6Fertig C and Gibble K 2000 Phys. Rev. Lett. 85 1622.
  • 7Guéna J, and Li R X, Gibble K, Bize S and Clairon A 2011 Phys. Rev. Lett. 106 130801.
  • 8Zhang S G 2004 Déplacement de fréquence d?au rayonnement du corps noir dans une fontaine atomique à césium et amélioration des performances de l'horloge.
  • 9Clairon A, Laurent P, Santarelli G, Ghezali S, Lea S N and Bahoura M 1995 Instrumentation and Measurement, IEEE Transactions on 44 128.
  • 10Shi C Y, Wei R, Zhou Z C, Li T, Li L and Wang Y Z 2011 Frequency Control and the European Frequency and Time Forum (FCS), 2011 Joint Conference of the IEEE International.

共引文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部