摘要 为了验证螺旋断层放射治疗(TOMO)机房的辐射防护设计方案的可靠性,根据相关规范计算螺旋断层放射治疗(TOMO)机房屏蔽体的理论计算结果,将辐射防护方案TOMO机房各侧屏蔽体设计厚度和机房屏蔽体的理论计算厚度进行比较。结果表明,运用该规范核算螺旋断层放射治疗(TOMO)机房各屏蔽体厚度的结果是可靠的,取值是合理的,机房屏蔽体理论计算厚度小于机房屏蔽体设计厚度,辐射防护设计方案的屏蔽效果满足国家现行标准的要求。 In order to verify the reliability of the radiation protection design scheme for the spiral tomotherapy(TOMO)machineroom,calculates the theoretical calculation results of the shielding body of the TOMO machine room according to relevant specifications,and compares the design thickness of the shielding body on each side of the TOMO machine room with the theoretical calculation thickness of the shielding body of the machine room.The results indicate that the calculation of the thickness of each shielding body in the spiral tomotherapy(TOMO)room using this specification is reliable and the values are reasonable.The theoretical calculated thickness of the shielding body in the room is less than the design thickness of the shielding body in the room,and the shielding effect of the radiation protection design scheme meets the requirements of current national standards.
作者 赵红 陶由之 Zhao Hong;Tao Youzhi(Sichuan Nuclear Industry Radiation Testing and Protection Institute(Sichuan Nuclear Emergency Technical Support Center),Chengdu,Sichuan 610052)
出处 《现代工程科技》 2023年第19期67-70,共4页 Modern Engineering Technology
关键词 TOMO 屏蔽计算 辐射防护 环境保护 TOMO shielding calculation radiation protection environmental protection
  • 相关文献

参考文献4

二级参考文献75

  • 1Mackie T R, Holmes T Swerdloff S,et al. Tomotherapy: A new concept for the delivery of dynamic eonformal radiotherapy [J]. Med Phys, i993,20(6) : 1 709-1 719.
  • 2Mackie T R, Holmes T W, Reckwerdt P J,et al.Tomotherapy: optimized planning and delivery of radiation therapy [J]. Int. J. Imaging. Sys. and Tech, 1995,6( 1 ) :43-55.
  • 3Yang J N, Mackie T R, Reckwerdt P J,et al. An investigation of Tomotherapy beam delivery [J]. Med Phys, 1997,24(3):425-436.
  • 4Mackie T R,Balog J,Ruchala K,et al. Tomotherapy[J]. Sere Radiat Oncol, 1999,9 ( 1 ) : 108-117.
  • 5Olivera G H,Shepard D M ,Ruchala K,et al.Tomotherapy Modern Technology of Radiation Oncologyed J Van Dyk [M]. Madison: Medical Physics Publishing, 1999:521-587.
  • 6Balog J P,Mackie T R,Reckwerdt P,et al.Characterization of the output for helical delivery of intensity modulated slit beams [J]. Med Phys, 1999,26 ( 1 ) : 55-64.
  • 7Balog J P,Mackie T R ,Wenman D L,et al.Multileaf collimator interleaf transmission [J]. Med Phys, 1999,26(2) : 176-186.
  • 8Balog J P,Mackie T R,Pearson D,et al.Benehmarking beam alignment for a clinical helical Tomotherapy device [J]. Med Phys, 2003, 30(6):1 118-1 127.
  • 9Balog J,Olivera G,Kapatoes J.Clinical helical Tomotherapy commissioning dosimetry [J]. Med Phys ,2003,30(12) :3 097-3 106.
  • 10Fenwick J D,Tome W A ,Jaradat H A,et ol. Quality assurance of a helical Tomotherapy machine [J]. Phys Med Biol,2004,49 (13):2 933-2 953.