摘要 AIM:To evaluate the efficacy and safety of Usights UC100 illuminated microcatheter in microcatheter-assisted trabeculotomy(MAT).METHODS:Totally 10 eyes of 10 patients with primary open angle glaucoma(POAG)who underwent MAT facilitated by Usights UC100(5 eyes)or iTrack(5 eyes)were reviewed.The success of this surgery was defined as intraocular pressure(IOP)<22 mm Hg with>30%reduction,without oral glaucoma medications,or additional glaucoma surgery.RESULTS:The mean pre-operative IOP was 25.38±10.22 mm Hg in the Usights UC100 group and 19.98±3.87 mm Hg in the iTrack group.MAT was achieved in all eyes in both groups.The success rates for the Usights UC100 group and iTrack groups were in all and 4 eyes,respectively.Both microcatheters produced a statistically significant reduction in IOP,and eyes using Usights UC100 achieved a lower IOP than the iTrack group at 3mo followup(12.58±1.52 and 14.84±1.89 mm Hg,respectively),but no statistical significance was there.No severe side effects were observed in either group.CONCLUSION:MAT using Usights UC100 or iTrack both achieve significant pressure reduction in cases of POAG,and Usights UC100 is as safe as iTrack.