真空式高温黑体辐射源性能评价

Performance evaluation of vacuum high⁃temperature blackbody radiation source
下载PDF
导出
摘要 为解决真空环境下辐射热流传感器的校准问题,采用热解石墨作为加热元件,通过标准光电高温计测量并控制温度,研建了50 mm大口径真空式高温黑体辐射源装置。开展了装置性能指标的测试试验和评价,结果显示装置10 min内的温度稳定性不超过0.5℃,径向温度均匀性不超过0.1%t(t为设定温度点),符合相应标准要求。最后计算分析了热流校准过程中真空式高温黑体辐射源装置性能指标引入的不确定度分量,为推动该装置在热流校准、辐射温度校准等领域中的应用提供了重要支撑。 In order to solve the calibration problem of radiation heat flux sensors in vacuum environments,a 50 mm large⁃diameter vacuum high⁃temperature blackbody radiation source device was developed by using pyrolytic graphite as the heating element,and measuring and controlling the temperature using a standard photoelectric pyrometer,and the performance indexes of the device were tested and evaluated.The temperature stability of the device within 10 min was not more than 0.5℃,and the radial temperature uniformity was not more than 0.1%t(t is the set temperature),which met the requirements of relevant regulations.Finally,the uncertainty components introduced by the performance indexes of the vacuum high⁃temperature blackbody radiation source in the process of heat flow calibration were calculated and analyzed.The research results provide strong support for promoting the application of the device in the calibrations of heat flow,ra⁃diation temperature and other fields.
作者 董磊 任佳 张岚 胡玮宸 DONG Lei;REN Jia;ZHANG Lan;HU Weichen(Changcheng Institute of Metrology&Measurement,Beijing 100095,China)
出处 《计测技术》 2023年第6期114-120,共7页 Metrology & Measurement Technology
基金 国家“十四五”计量技术基础科研项目(JSJL2020205B003)。
关键词 高温黑体 热流 温度校准 high⁃temperature blackbody heat flux temperature calibration
  • 相关文献

参考文献9

二级参考文献68

  • 1王铁军,柏成玉,原遵东,董伟,王景辉,段宇宁.高温共晶固定点的建立与复现[J].计量学报,2007,28(z1):23-27. 被引量:5
  • 2罗拥军.钨铼热电偶及其应用[J].仪器仪表用户,1994,7(4):6-9. 被引量:2
  • 3王学新,岳文龙,杨鸿儒,谢毅.IRS400型材料发射率测试装置的研制[J].应用光学,2015,36(2):272-276. 被引量:3
  • 4黄岳巍,崔瑞祯,巩马理,陈刚.基于TEC的大功率LD恒温控制系统的研究[J].红外与激光工程,2006,35(2):143-147. 被引量:30
  • 5行联合,曹仙春,杨平.半导体热电温控模块的设计应用[J].机电工程技术,2007,36(8):84-87. 被引量:1
  • 6WOOLLIAMS E R, FOX N P, COX M G, et al. The CCPR Kl-a key comparison of spectral irradi- ance from 250 nm to 2 500 nm: measurements analy-sis and results[J].Metrology, 2006, 43:S98-S104.
  • 7KLEIN R, TAUBERTt D, THORNAGEL R, et al. Radiometric comparison of the primary synchrotron radiation source standard metrology light source with calibrated filter radiometers in the visible and NIR spectral range[J]. Metrology, 2009, 46:359-366.
  • 8SHAW P S, ARP U, SAUNDERS R D, et al. Syn- chrotron radiation-based irradiance calibration from 200 to 400 nm at the synchrotron ultraviolet radiation facility III[J]. Applied Optics, 2007, 46:25-35.
  • 9RICHTER M, HOLLANDT J, KROTH U, et al. Source and detector calibration in the UV and VUV at bessy II[J]. Metrology, 2003, 40:S107-S110.
  • 10ZWINKELS J C, IKONEN E, FOX N P, et al. Photometry, radiometry and ' the candela': evolu- tion in the classical and quantum world[J]. Metrolo- gy, 2010, 47 :R15-R32.

共引文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部