50~120kV脉冲X射线参考辐射场建立及剂量率测量

Establishment of 50-120 kV Pulsed X-ray Reference Radiation Field and Dose Rate Measurement
下载PDF
导出
摘要 为满足国内医疗、安检等脉冲辐射场所对剂量监测仪表准确溯源的需求,基于栅控技术建立满足规范和标准要求的50~120 kV毫秒级脉冲X射线参考辐射场,其脉冲宽度在1~6 300 ms范围内可调。通过对该辐射场的管电压稳定性、剂量线性、重复性、脉冲波形等关键参数进行研究测试,建立脉冲X射线参考波形,并使用30 cm3的圆柱电离室开展脉冲剂量率测量研究。结果表明,管电压预设60 kV时,介入法、非介入法测量管电压的相对偏差均不超过±1.5%;可输出参考波形脉冲的最小脉冲宽度为20 ms,在1 m处95%均匀辐射野直径约为5.5 cm;且N系列、RQR系列辐射质条件下的剂量率范围可分别扩展至0.002~10 mGy/s和0.03~250 mGy/s,可用于防护仪表的量值溯源。 In order to meet the needs of accurate traceability of dose monitoring instruments in domestic medical,security inspection and other pulse radiation places,a 50-120 kV millisecond pulse X-ray reference radiation field that meets the requirements of norms and standards was established,based on the gate control technology,and its pulse width can be adjustable in the range of 1-6300 ms.The pulse X-ray reference waveform was established by studying and testing the tube voltage stability,dose linearity,repeatability,pulse waveform and other key parameters of the radiation field,and the pulse dose rate was measured using a cylindrical ionization chamber with a sensitive volume of 30 cm 3,and obtained the following research results:when the tube voltage was preset to 60 kV,the relative deviation of the intervened and non-intervened measuring tube voltage did not exceed±1.5%;the minimum pulse width of the reference waveform pulse output was 20 ms,and a 95%uniform radiation field with a diameter of about 5.5 cm could be generated at 1 m.Moreover,the dose rate range of N series and RQR series radiation quality was extended to 0.002-10 mGy/s and 0.03-250 mGy/s,respectively,which can be used for the metrological traceability of protective instruments.
作者 郝光辉 张晓乐 徐坠 李德红 黄建微 成建波 万琳健 HAO Guanghui;ZHANG Xiaole;XU Zhui;LI Dehong;HUANG Jianwei;CHENG Jianbo;WAN Linjian(National Institute of Metrology,Beijing 100029,China)
出处 《同位素》 CAS 2024年第1期7-16,共10页 Journal of Isotopes
基金 国家重点研发计划NQI项目(2017YFF0206205)。
关键词 脉冲X射线 参考辐射场 参考脉冲波形 剂量率测量 辐射剂量学 pulse X-ray reference radiation field reference pulse waveform dose rate measurement radiation dosimetry
  • 相关文献

参考文献4

二级参考文献20

  • 1中华人民共和国国家质量监督检验检疫总局,GB18871-2002电离辐射防护与辐射源安全基本标准[S].北京:中国标准出版社,2002
  • 2Anagnostopoulos G, Bahas D, Karaiskos P, et al. Thermoluminescent dosimetry of the selectseed 125][ interstitial brachytherapy seed [ J]. Med Phys, 2002, 29 (5) : 709-716.
  • 3Lymperopoulou G, Papagiannis P, Sakelliou L, et al. Monte Carlo and thermoluminescence dosimetry of the new IsoSeed model I25. S17 125I interstitial brachytherapy seed [ J]. Med Phys, 2005, 32 ( 11 ) : 3313-3317.
  • 4Nath R, Anderson LL, Luxton G, et al. Dosimetry of interstitial brachytherapy sources: recommendations of the AAPM Radiation Therapy Committee Task Group No 43[J]. Med Phys, 1995, 22 ( 2 ) : 209-234.
  • 5Rivard M J, Coursey BM, Dewerd LA, et al. Update of AAPM Task Group No 43 report: a revised AAPM protocol for brachytherapy dose calculations[ J]. Med Phys, 2004, 31 ( 3 ) : 633 -674.
  • 6Tang K, Cui H, Zhu H, et al. Study of a new Lif: Mg, Cu, P formulation with enhanced thermal stability and a lower residual TL signal[J]. Radiat Meas, 2007, 42(1): 24-28.
  • 7Olko P, Bilski P, Ryba E, et al. Microdosimetric interpretation of the anomalous photon energy response of ultra-sensitive LiF: Mg, Cu, PTLdosemeters[J]. RadiatProtDosim, 1993, 47(1- 4) : 31-35.
  • 8Fill UA, Regulla DF. Advanced LiF technology for the assessment of patient exposure in diagnostic radiology [ J ]. Radiat Prot Dosirn, 1998, 80 ( 1-3 ) : 225-229.
  • 9Konnai A, Nariyama N, Ohnishi S, et al. Energy response of LiF and Mg2SiO4 TLDs to 10-150 keV rnonoenergetic photons [J]. Radiat Prot Dosirn, 2005, 115(1-4): 334-336.
  • 10Massillon-J LG, Cabrera-Santiago A, Minniti R, et al. Influence of phantom materials on the energy dependence of LiF: Mg, Ti thermolurninescent dosimeters exposed to 20-300 kV narrow X-ray spectra, 137Cs and 6Co photons[J]. Phys Med Biol, 2014, 59 (15) : 4149-4166.

共引文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部