呼吸机常规复用管路与一次性管路的质控研究

Research on the Quality Control of Routine Reusable Pipeline and Disposable Pipeline of Ventilator
下载PDF
导出
摘要 目的探讨常规复用管路和一次性管路对呼吸机质控结果的影响。方法随机抽取17台呼吸机,分别采用常规复用管路和一次性管路进行质控,记录质控数据,然后采用配对t检验分析2种管路的差异是否显著。结果对于呼吸频率、潮气量和气道末压,2种管路没有显著差异(P>0.05);对于气道峰压,常规复用管路显著高于一次性管路(P<0.05),但差异非常小,仅0.2 mbar(1 mbar=100 Pa)左右,不影响质控结论。结论常规复用管路和一次性管路不影响呼吸机质控结果,2种管路可以相互替代。 Objective To explore the effect of routine reusable pipeline and disposable pipeline on ventilator quality control results.Methods 17 ventilators were randomly selected to conduct quality control using routine reusable pipeline and disposable pipeline respectively.Quality control data were recorded and then paired t-test method was used to analyze whether the difference between the two pipelines was significant or not.Results There were no significant differences in respiratory rate,tidal volume and end airway pressure between the two types of pipes(P>0.05).The airway peak pressure of routine reusable pipeline was significantly higher than disposable pipeline(P<0.05),but the difference was very small,only about 0.2 mbar which would not affect the conclusion of quality control.Conclusion Quality control of ventilator is not affected by routine reusable pipeline and disposable pipeline,which can be replaced by each other.
作者 鲍莉 沈云明 向思伟 郑焜 BAO Li;SHEN Yunming;XIANG Siwei;ZHENG Kun(Children’s Hospital,Zhejiang University School of Medicine,Hangzhou,310057)
出处 《中国医疗器械杂志》 2024年第1期111-113,共3页 Chinese Journal of Medical Instrumentation
关键词 呼吸机 常规复用管路 一次性管路 质控 ventilator routine reusable pipeline disposable pipeline quality control
  • 相关文献

参考文献9

二级参考文献74

共引文献38

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部