可吸收支架负载外泌体方法研究

Research on the method of loading exosomes onto absorbable stents
原文传递
导出
摘要 目的探讨一种将外泌体负载于可吸收支架的方法。方法应用化学连接法,通过建立支架-氨丙基三乙氧基硅烷-二硬脂酰磷脂酰乙酰胺-聚乙二醇5000-外泌体的连接使可吸收支架负载外泌体,构建外泌体洗脱可吸收支架。扫描电镜观察支架表面及外泌体吸收情况,飞行时间质谱对支架表面物质进行鉴定,原子力显微镜检测支架表面粗糙程度,体视显微镜观察支架外观及外径,径向力检测仪检测支架力学强度;以单纯可吸收支架作为对照。结果扫描电镜观察示外泌体洗脱可吸收支架表面有细小不规则裂纹以及许多外泌体颗粒。原子力显微镜示5μm2取样范围其表面粗糙程度为±70 nm,而可吸收支架为±20 nm。飞行时间质谱检测示外泌体洗脱可吸收支架与外泌体在质荷比(mass charge ratio,m/z)处均有峰值,而可吸收支架无此峰值;并且外泌体洗脱可吸收支架m/z73处峰值较可吸收支架明显降低。体视显微镜观察示外泌体洗脱可吸收支架外径在可吸收支架规格范围内且表面无明显裂纹、毛刺及凹陷;径向强度检测结果符合原可吸收支架径向强度标准。结论应用化学连接法可以将外泌体负载于可吸收支架表面,并且支架尺寸、径向强度均符合原可吸收支架标准。 Objective To explore a method of loading exosomes onto absorbable stents.Methods By building a stent-(3-aminopropyl)triethoxysilane-1,2-distearoyl-sn-glycero-3-phosphoethanolamine-N-[methoxy(polyethylene glycol)5000]-exosomes connection,the exosomes were loaded onto absorbable stents to obtained the exosome-eluting absorbable stents.The surface conditions of the stents and absorption of exosomes were observed by scanning electron microscope and identified through the time-of-flight mass spectrometry;the roughness of the stents’surfaces was observed by atomic force microscope;the appearances and sizes of the stents were observed by stereomicroscope;and the radial force was tested by tensile test machine.The absorbable stents were used as control.Results The scanning electron microscope observation showed that the exosome-eluting absorbable stents had some small irregular cracks on the surface where many exosomes could be seen.The atomic force microscopy observation showed that within the range of 5μm2,the surface roughness of the absorbable stents was±20 nm,while the surface roughness of the exosome-eluting absorbable stents was±70 nm.In the results of time-of-flight mass spectrometry,both the exosome-eluting absorbable stents and exosomes had a peak at the mass charge ratio of 81(m/z 81),while the absorbable stents did not have this peak.The peak of exosome-eluting absorbable stents at m/z 73 showed a significant decrease compared to the absorbable stents.The stereomicroscope observation showed that the sizes of exosome-eluting absorbable stents met standards and the surfaces had no cracks,burrs,or depressions.The radial force results of the exosome-eluting absorbable stents met the strength standards of the original absorbable stent.Conclusion By applying the chemical connection method,the exosomes successfully loaded onto the absorbable stents.And the sizes and radial forces of this exosome-eluting absorbable stents meet the standards of the original absorbable stents.
作者 王宸宸 李天博 郭瑞敏 季庆辉 王江宁 曹毓琳 WANG Chenchen;LI Tianbo;GUO Ruimin;JI Qinghui;WANG Jiangning;CAO Yulin(Jiamusi University,Jiamusi Heilongjiang,154007,P.R.China;Department of Orthopedics,Shijitan Hospital Affiliated to Capital Medical University,Beijing,100038,P.R.China;Tangyi Holdings(Shenzhen)Co.,Ltd,Shenzhen Guangdong,518000,P.R.China)
出处 《中国修复重建外科杂志》 CAS CSCD 北大核心 2024年第2期206-210,共5页 Chinese Journal of Reparative and Reconstructive Surgery
基金 北京炎华血管与创面诊疗技术发展基金会资助项目(2022-B01) 黑龙江省省属高等学校基本科研业务费基础研究项目(2021-KYYWF-0604)。
关键词 外泌体 可吸收支架 表征 Exosome absorbable stent characterization
  • 相关文献

参考文献6

二级参考文献48

共引文献159

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部