摘要
使用MCNP5对两个高气压电离室进行MC模拟,并完成实验测量,分别得到两个电离室的两种能量响应结果。将模拟结果与实验结果对比发现,两种结果在低能量段差异较大,两个电离室都在48 keV处相对偏差最大;在高能量段差异较小,1号电离室在1250 keV处相对偏差最大,2号电离室在662 keV处相对偏差最大;两种结果在164~1250 keV能量段的能量响应整体范围相近。最后分析发现在低能量段差异较大的主要原因是电离室的制造误差和模拟计算能量沉积时产生的计算误差。
MCNP5 is used to simulate the two high pressure ionization chambers,and the experimental measurements are completed.Two kinds of energy response results of the two ionization chambers are obtained respectively.Comparing the simulation results with the experimental results,it is found that the two results differ greatly in the low energy range,and the relative deviation of the two ionization chambers is the largest at 48 keV;In the high energy range,the difference is small.The relative deviation of No.1 ionization chamber is the largest at 1250 keV,and that of No.2 ionization chamber is the largest at662 keV;The energy response of the two results in the energy range of 164 keV to 1250 keV is similar.Finally,it is found that the main reasons for the large difference in the low energy range are the manufacturing error of the ionization chamber and the calculation error generated in the simulation calculation of energy deposition.
作者
漆成龙
丁卫撑
黄平
龚岚
刘博帅
QI Chenglong;DING Weicheng;HUANG Ping;GONG Lan;LIU Boshuai(College of Nuclear Technology and Automation Engineering,Chengdu University of technology,Chengdu 610059,China;Division of Ionizing Radiation Metrology of National Institute of Measurement and Testing Technology,Chengdu 610021,China)
出处
《核电子学与探测技术》
CAS
北大核心
2023年第6期1126-1130,共5页
Nuclear Electronics & Detection Technology
基金
国家自然科学基金(41774140)
成都市重大科技创新项目(021-YF08-00046-GX)。
关键词
MCNP5
高气压电离室
MC模拟
实验测量
能量响应
计算误差
MCNP5
high pressure ionization chamber
Monte Carlo simulation
experimental measurement
energy correspondence
calculation error