期刊文献+

BacT/ALERT两种血培养系统8种血培养瓶对模拟菌血症标本检测能力的对比研究

Comparative study on the detection ability of BacT/ALERT two blood culture systems and eight blood culture bottles for simulated bacteremia specimens
原文传递
导出
摘要 目的比较需氧和兼性厌氧血培养瓶BacT/ALERT FA Plus、FA、PF Plus、PF、SA及厌氧和兼性厌氧血培养瓶BacT/ALERT FN Plus、FN、SN对模拟菌血症标本检测的差异;验证BacT/ALERT VIRTUO血培养仪监测血量的准确性。方法选取23种常见菌配制成菌悬液,在培养瓶中加入定量无菌缓冲液和健康人血以及菌悬液,分别放入两个血培养仪进行培养,比较平均检测时间(TTD)和阳性生长情况;向血培养瓶中注入定量磷酸缓冲盐溶液(PBS)后放入BacT/ALERT VIRTUO血培养仪,记录仪器监测量,并计算其与注入量的差异。结果除人型葡萄球菌和新型隐球菌外,8种血培养瓶对其余21种菌的阳性检出率均为100%;5种需氧和兼性厌氧瓶检测需氧和兼性厌氧微生物的平均TTD差异无统计学意义(P>0.05);在检测厌氧和兼性厌氧菌方面,SN最快。在检测需氧和兼性厌氧菌、厌氧和兼性厌氧菌方面,BacT/ALERT VIRTUO血培养仪比国内使用较广泛的BacT/ALERT 3D血培养仪平均TTD分别缩短1.5~8 h和3~8 h;在检测真菌方面,两仪器间差异无统计学意义(P>0.05)。BacT/ALERT VIRTUO仪器监测血量和实际注入量之间差异无统计学意义(P>0.05)。结论5种需氧和兼性厌氧瓶在检测常见需氧和兼性厌氧菌血症标本方面的能力差异无统计学意义,SN在缩短厌氧和兼性厌氧菌血症检测时间方面优于FN和FN Plus;BacT/ALERT VIRTUO血培养仪能明显缩短菌血症标本的TTD,且在监测血量方面准确可靠。 Objective To compared the difference of aerobic and facultative anaerobic blood culture bottles BacT/ALERT FA Plus,FA,PF Plus,PF and SA,as well as anaerobic and facultative anaerobic bottles BacT/ALERT FN Plus,FN and SN in the detection of simulated bacteremia samples,and to verify the accuracy of monitoring blood volume with BacT/ALERT VIRTUO blood culture instrument.Methods 23 kinds of bacteria were selected to prepare bacterial suspension.Healthy human blood and bacterial suspension were added into the bottles,and cultured in two blood culture systems respectively.The mean detection time(time-to-detect,TTD)and positive growth were compared.A certain amount of phosphate buffer saline(PBS)was injected into the blood culture bottles,and then placed in the BacT/ALERTVIRTUO blood culture instrument to get the monitoring volume,and calculate the difference between the injected volume and the monitoring volume.Results Except for Staphylococcus hominis and Cryptococcus neoformans,the positive detection rates of the other 21 strains were 100%.There was no significant difference in the average TTD time of the five kinds of aerobic bottles in the detection of aerobic and facultative anaerobic bacteria and fungi.For the detection of anaerobes and facultative anaerobes,SN bottle was the fastest.In the detection of aerobic and facultative anaerobes,anaerobic and facultative anaerobes,the average TTD time of BacT/ALERTVIRTUO blood culture instrument was shortened by 1.5~8 h and 3~8 h compared with BacT/ALERT 3D blood culture instrument,and the difference was statistically significant.The difference between the monitoring volume and the actual injection volume was no statistical difference.Conclusions There was no significant difference in the ability of five kinds of aerobic bottles to detect common bacteremia specimens.The anaerobic bottle SN is superior to FN and FN Plus bottles in shortening the detection time.BacT/ALERTVIRTUO blood culture instrument can significantly shorten the detection time of common bacteremia samples and is accurate and reliable in monitoring blood volume.
作者 王晓苏 戴铮 朱嘉嘉 李启超 张李涛 Wang Xiaosu;Dai Zheng;Zhu Jiajia;Li Qichao;Zhang Litao(Department of Clinical Laboratory,Wuhan Asia General Hospital,Wuhan Hubei 430050,China)
出处 《中华临床实验室管理电子杂志》 2023年第4期207-213,共7页 Chinese Journal of Clinical Laboratory Management(Electronic Edition)
基金 武汉市医学科研项目(WX21D49)。
关键词 血培养 培养基 细菌 血流感染 blood culture culture medium bacteria bloodstream infection
  • 相关文献

参考文献2

二级参考文献14

  • 1Egidio FV,Emanuela V,Carlo A,et al.Use of simulated blood cultures for time to detection comparison between BacT/ALERTTM and BACTECTM 9240 blood culture syBtems.Diagn Micmbiol Infect Dis,2002,44:235-240.
  • 2Zieger R,Johnscher I,Martus P,et al.Controlled clinical laboratory compareison of two supplemented aerobic and anaerobic media used in automated blood culture systems to detect bloodstream infections.J Clin Microbiol,1998,36:657-661.
  • 3Lynn LH,Bebjamin JG,Clinton KM,et al.Direct comparison of the BACTEC 9240 and BacT/ALERT 3D automated blood culture systems for candida growth detection.J Clin Micrebiol,2004,42:115-118.
  • 4Spaargaren J,Boven CP,Voorn GP.Effectiveness of resins in neutralizing antibiotic activities in Bactec Plus Aerobic/F culture medium.J Clin Microbiol,1998,36:3731-3733.
  • 5Stanley M,Reller LB,Petti CA.Controlled Clinical Comparison of BacT/ALERT Standard Aerobic medium with BACTEC Standard Aerobic Medium for Culturing Blood.J Clin Mierobiol,2003,41:2391-2394.
  • 6Ozay AA,Eda Y.Comparison of the effect of delayed entry into 2 different blood culture systems(BACTEC 9240 and BacT/ALERT 3D)on culture positivity.Diagn Micrebiol Infect Dis,2006,54:193-196.
  • 7Grog SM,David MM,Stephanie E,et al.The epidemiology of sepsis in the United States from 1979 through 2000.N Engl J Med,2003,348:1546—1554.
  • 8Rangel—Frausto MS,Pittet D,Cosfigan M,et al.The natural history of the systemic inflammatory response syndrome(SIRS),a prospective study.JAMA,1995,273:117—123.
  • 9Viganò EF,Vasconi E,Agrappi C,et al.Use of simulated blood cultures for antibiotic effect on time to detection of the two blood culture systems BacT/ALERT and BACTEC 9240.New Microbiol,2004,27:235-248.
  • 10Flayhart D,Borek A,Wakefield T,et al.Ability of BD BACTEC Plus blood culture bottles versus BacT/Alert FAN blood culture bottles to detect bacterial pathogens samples containing vacomycin.The 105^th General Meeting of the American Society for Microbiology.2005.