期刊文献+

瓣叶图案化对机械瓣膜流体力学性能的影响

Effect of Leaflet Pattering on Mechanical Valve Hydrodynamic Properties
下载PDF
导出
摘要 目的探讨在机械瓣瓣叶表面构建图案化织构是否对机械瓣流体力学性能产生影响。方法分别随机抽取GKS 21A、23A两种规格的双叶主动脉机械瓣,作为对照组按照ISO 5840-2:2021标准要求用脉动流测试仪测试瓣膜的平均跨瓣压差(mean transvalvular pressure,MTP)、反流百分比(regurgitation fraction,RE)、有效开口面积(effective orifice area,EOA)、能量损耗(energy loss,eLoss)等流体力学指标;然后用激光刻蚀的方法在每个瓣叶表面构建垂直于血流方向的横纹图案,再次对该实验组瓣膜进行同样的测试,记录、分析所得数据。结果在脉动流试验中,实验组的MTP在相同流量、规格下比对照组小;GKS 21A实验组瓣膜的RE比对照组小,GKS 23A实验组瓣膜的RE比对照组大;实验组瓣膜的EOA在相同流量、规格下均比对照组大;GKS 21A实验组瓣膜的eLoss比对照组小,GKS 23A实验组瓣膜的eLoss比对照组大。结论机械瓣瓣叶表面横纹图案会影响到瓣膜的血流动力学性能。 Objective To investigate the effects patterned structure located on the surface of mechanical valve leaflet exerted on the hydrodynamic properties of valve prostheses.Methods Bileaflet aortic mechanical valves of GKS 21A and 23A were randomly selected as the control group,hydrodynamic performance of each valve were assessed,as per ISO5840 by calculating mean transvalvular pressure(MTP),regurgitation fraction(REG),effective orifice area(EOA)and effective orifice area(eLoss).Then,laser etching was applied to construct a parallel-groove array pattern on leaflet surface perpendicular to the direction of the blood flow.The valve was subjected again to the same test,and the data obtained were recorded and analyzed.Results In the pulsating flow test,the MTP of the experimental group was smaller than that of the control group at the same flow rate and size.The RE in GKS 21A experimental group was smaller than that in control group,and the RE in GKS 23A experimental group was larger than that in control group.The EOA in the experimental group was larger than that in the control group at the same flow rate and size.The eLoss in the GKS 21A experimental group was smaller than that in the control group,and the eLoss in the GKS 23A experimental group was larger than that in the control group.Conclusion The parallel-groove array pattern on the surface of the leaflet affected the hemodynamic performance of the valve prostheses.
作者 王玉苗 周建业 刘丽 WANG Yumiao;ZHOU Jianye;LIU Li(State Key Laboratory of Cardiovascular Disease,National Center for Cardiovascular Diseases,Fuwai Hospital,Peking Union Medical College and Chinese Academy of Medical Sciences,Beijing 100037,China)
出处 《医学研究杂志》 2024年第2期36-39,46,共5页 Journal of Medical Research
基金 国家重点研发计划项目(2018YFC1105502)。
关键词 表面图案化 血流动力学 脉动流测试仪 人工心脏瓣膜 Patterns of leaflets Hemodynamics Pulse duplicator Heart valve prosthesis
  • 相关文献

参考文献2

二级参考文献7

共引文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部