摘要
针对超声骨刀在手术过程中切割效率低和产生骨组织热损伤的问题,设计了具有二级放大结构的高载能超声骨刀。基于驻波理论的等效长度法,结合有限元仿真对其二级放大结构进行设计和优化,测试了其振动性能并进行了骨切割试验。结果表明:具有二级放大结构的超声骨刀的谐振频率为23306 Hz,振幅为140μm,骨切割表面平整且无损伤,满足临床手术需求。
Aiming at the problems of low cutting efficiency of ultrasonic bone scalpel and the generation of thermal damage of bone tissue during surgery,a high-capacity ultrasonic bone scalpel with a secondary amplification structure was designed.Based on the equivalent length method of standing wave theory and finite element simulation,the design and optimization of the secondary amplified structure were carried out,and the vibration performance was tested and the bone cutting test was conducted.The results show that the resonance frequency of the ultrasonic bone scalpel with secondary amplification structure is 23306 Hz,the amplitude is 140μm,and the bone cutting surface is flat and damage-free,which meets the needs of clinical surgery.
作者
姜兴刚
邓康
尹欣
江培源
JIANG Xinggang;DENG Kang;YIN Xin;JIANG Peiyuan(School of Mechanical Engineering and Automation,Beihang University,Beijing 100191,China;Senior Department of Orthopedics,The Fourth Medical Center,Chinese People's Liberation Army General Hospital,Beijing 100048,China;Aero Engine Corporation of China Guiyang Engine Design Research Institute,Guiyang 550081,Guizhou,China)
出处
《陕西师范大学学报(自然科学版)》
CAS
CSCD
北大核心
2024年第2期1-7,共7页
Journal of Shaanxi Normal University:Natural Science Edition
基金
国家自然科学基金(51975035)。
关键词
超声骨刀
骨组织
二级放大
驻波理论
有限元仿真
ultrasonic bone scaple
bone tissue
secondary amplification
standing wave theory
finite element simulation