期刊文献+

基于光度滴定法研制的COD_(Mn)在线分析仪

Development of COD_(Mn) Online Analyzer Based on photometric Titration
下载PDF
导出
摘要 基于光度滴定法研制了一种高锰酸盐指数在线分析仪,介绍了分析仪的分析流路,滴定反应装置,并对滴定终点判定方法进行了优化,同时对分析仪的性能进行了测定。研究结果表明:在0~20mg/L的量程范围内,分析仪的标准曲线的R^(2)值为0.9999;定量下限为0.29mg/L;重复性误差为0.42%,零点漂移与量程漂移分别为0.31%与0.35%,葡萄糖试验为0.61%,实际水样比对试验为6.92%。本分析仪方法灵敏度高,准确度高,精密度好,定量下限低,适于现场水质监测。 A permanganate index online analyzer is developed based on the luminosity titration method,and the titration endpoint determination method is analyzed,and the performance of the analyzer is determined.In the range of 0~20 mg/L,the results show that the R^(2) value of the standard curve of the analyzer is 0.9999;the lower quantitative limit is 0.29 mg/L;the repeatability error was 0.42%,the zero drift and range drift were 0.31%and 0.35%,respectively;the glucose test was 0.61%,and the actual water sample comparison test was 6.92%.This analyzer method has high sensitivity,high accuracy,good precision,low lower limit of quantification,and is suitable for site water quality monitoring.
作者 樊小燕 朱志芳 王超 洪陵成 Fan Xiaoyan;Zhu Zhifang;Wang Chao;Hong Lingcheng(Jiangsu Delin Environmental Protection Technology Co.,LTD.,Nanjing 211103,China;Nanjing nmol Instrument Co.,Ltd,Nanjing 211103,China)
出处 《现代科学仪器》 2024年第1期16-19,31,共5页 Modern Scientific Instruments
基金 港澳台科技合作项目(No.BZ2021056)资助。
关键词 光度滴定法 滴定终点判定算法 高锰酸盐指数 在线分析仪 photometric titration method titration end point determination algorithm permanganate index online analyzer
  • 相关文献

参考文献1

二级参考文献4

共引文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部