摘要
●AIM:To quantify the performance of artificial intelligence(AI)in detecting glaucoma with spectral-domain optical coherence tomography(SD-OCT)images.●METHODS:Electronic databases including PubMed,Embase,Scopus,ScienceDirect,ProQuest and Cochrane Library were searched before May 31,2023 which adopted AI for glaucoma detection with SD-OCT images.All pieces of the literature were screened and extracted by two investigators.Meta-analysis,Meta-regression,subgroup,and publication of bias were conducted by Stata16.0.The risk of bias assessment was performed in Revman5.4 using the QUADAS-2 tool.●RESULTS:Twenty studies and 51 models were selected for systematic review and Meta-analysis.The pooled sensitivity and specificity were 0.91(95%CI:0.86–0.94,I2=94.67%),0.90(95%CI:0.87–0.92,I2=89.24%).The pooled positive likelihood ratio(PLR)and negative likelihood ratio(NLR)were 8.79(95%CI:6.93–11.15,I2=89.31%)and 0.11(95%CI:0.07–0.16,I2=95.25%).The pooled diagnostic odds ratio(DOR)and area under curve(AUC)were 83.58(95%CI:47.15–148.15,I2=100%)and 0.95(95%CI:0.93–0.97).There was no threshold effect(Spearman correlation coefficient=0.22,P>0.05).●CONCLUSION:There is a high accuracy for the detection of glaucoma with AI with SD-OCT images.The application of AI-based algorithms allows together with“doctor+artificial intelligence”to improve the diagnosis of glaucoma.