期刊文献+

一种用于多通道移液装置性能验证的光度法系统

A Photometry System for Performance Verification of Multichannel Liquid Delivery Devices
下载PDF
导出
摘要 针对多通道移液装置的校准和验证,本文介绍了一种全新的多通道验证系统(MVS),并基于双染料比率光度法,对每个单独通道的准确性和精密度进行测量。实验证明:该系统应用范围较广,校准范围可覆盖(2~200)μL,且结果可溯源至国家标准。同时,基于MVS自身的稳定性和操作的便利性,在实验室环境中可用于对自动移液设备进行验证,确保仪器设备的性能状态。 This paper introduces a new multi-channel verification system(MVS)for the calibration and verification of multi-channel pipettes,and measures the accuracy and precision of each individual channel based on double dye ratio photometry.Experiments show that the system has a wide application range,the calibration range can cover(2~200)μL,and the results can be traced to the national standard.At the same time,based on the stability of MVS itself and the convenience of operation,it can be used to verify the automatic pipetting equipment in the laboratory environment to ensure the performance status of the instrument.
作者 许诚 吴桂锋 雷鸣 韩可 李大伟 XU Cheng;WU Guifeng;LEI Ming;HAN Ke;LI Dawei
出处 《计量与测试技术》 2024年第2期37-40,44,共5页 Metrology & Measurement Technique
基金 2022年教育部高等学校科学研究发展中心中国高校产学研创新基金项目(项目编号:2022BL082) 2022年江苏省高等教育学会高校实验室研究委员会研究课题(项目编号:GS2022BZZ36) 2021年度江苏省高等学校基础科学(自然科学)研究面上项目(项目编号:21KJD460006)。
关键词 多通道移液器 多通道验证系统 验证与校准 体积测量 multichannel liquid delivery multichannel verification system verification and calibration volume measurement
  • 相关文献

参考文献5

二级参考文献19

  • 1丁继敏,范维珂.血清铁蛋白与恶性肿瘤[J].国外医学(临床生物化学与检验学分册),1993,14(4):152-154. 被引量:2
  • 2亓同钢,汪运山,刘相东,孟月生,张伟,关广聚.抑制AFP基因表达能直接抑制肝癌细胞系SMMC-7721增殖[J].中国肿瘤生物治疗杂志,2005,12(1):65-66. 被引量:1
  • 3McMillan N, O'Neill, M, Smith S, et al. Reliable TLDA-microvolume UV spectroscopy with applications in chemistry and biosciences for microlitre analysis and rapid pipette calibration I J]. Journal of Optics A: Pure and Applied Optics,2009, 11(5):1 -10.
  • 4Batista E, Arias R, Wang J T. Final report on BIPM/ CIPM key comparison CCM. FF-K4. 2. 2011: Volume comparison at 100 ixL--Calibration of micropipettes ( piston pipettes ) [ J ]. Metrologia, 2013,50 : 07003.
  • 5Batista E, Pinto L, Filipe E, et al. Calibration of micropipettes : Test methods and uncertainty analysis [ J ]. Measurement: Journal of the International Measurement Confederation,2007, 40 ( 3 ) : 338 - 342.
  • 6Aibe V Y. Water mass measurement m parts for volume standard calibration in tlie key comparison [ C ]// FLOMEKO, 14th International Flow Measurement Conference, Sandton, South Africa, 2007 : 145 - 148.
  • 7Madee T, Amarouehe S. Comparison of mieropipettes calibration by accredited laboratories [ C ]//3rd International Metrology Conference 2010, CAFMET 2010, 2010, Marrakech, Morocco,2010:337 - 343.
  • 8杨禹哲,廖鹏.农药残毒快速测定仪校准用标准物质及方法的研究[J].计量学报,2010,31(5A):78-79.
  • 9Batista E, Richard P. The selection of water property formulae for volume and flow calibration [ J ]. Metrologia, 2007, 44 (6) : 453 - 463.
  • 10Tanaka M, Girard G, Davis R, et al. Recommended table for the density of water between 0C and 40C based on recent experimental reports [ J ]. Metrologia, 2001,38: 301 -309.