期刊文献+

激光粒度分析仪测试条件对粒径分析的影响

The Effect of Test Conditions on the Particle Size Distribution Analysis in the Laser Particle Size Analyzer
下载PDF
导出
摘要 合适的激光粒度分析仪测试条件对于粒径分布数据的准确性至关重要。对比不同粒径粉末在欧美克LS-909激光粒度分析仪、SCF-105B湿法进样器中的测试结果并进行分析。结果表明,对于小于1μm的粉末样品,建议超声60 min以上,遮光率选择5%~8%,泵速选择2000~3000 r/min;对于几十μm的粉末样品,可不超声,遮光率选择10%~15%,泵速选择1500~2500 r/min;对于几百μm的样品,建议不超声,遮光率选择10%~20%,泵速选择2500~3500 r/min。在以上条件下测得的数据可靠且重复性高。 Appropriate testing conditions of laser particle size analyzer are very important for the accuracy of particle size distribution data.Powders with different particle sizes were tested under different test conditions,based on the OMEC LS-909 laser particle size analyzer and SCF-105B wet sample injector.It was found that the sample with particle size less than 1μm,is recommended to ultrasound for more than 60 minutes,with a shading rate of 5%~8%,and a pump speed of 2000~3000 r/min.However,for the sample with several tens of micrometers,ultrasound is not necessary.An appropriate testing condition is the shading rate of 10%~15%with pump speed of 1500~2500 r/min.For the sample with a few hundred microns,ultrasound is not necessary,the testing condition should be a shading rate of 10%~20%,and a pump speed of 2500~3500 r/min.The test results are reliable and repeatable under the above conditions.
作者 李丹 林丽萍 王雅 杨贵东 LI Dan;LIN Liping;WANG Ya;YANG Guidong(Instrumental Analysis Center,School of Chemical Engineering and Technology,Xi’an Jiaotong University,Xi’an 712046,China)
出处 《实验室研究与探索》 CAS 北大核心 2024年第1期32-35,共4页 Research and Exploration In Laboratory
关键词 公共分析测试 激光粒度分析仪 精密仪器 粒径 public analysis testing laser particle size analyzer precision instrument particle size
  • 相关文献

参考文献13

二级参考文献108

共引文献132

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部