摘要 Thiacloprid is the first-generation neonicotinoid insecticide,which is widely used in modern agriculture.Reference materials are essential for sound measurement and traceability of results in chemical analysis.Pure thiacloprid was comprehensively assessed by two orthonormal methods:mass balance(MB) and quantitative nuclear magnetic resonance spectroscopy(qNMR).From MB,its mass fraction value was 998.4 mg/g with the subtraction of structurally related impurities(99.87%),inorganic impurities(0.006 mg/g),water(0.33 mg/g),and volatile organic solvent(0 mg/g).To guarantee sufficient separation of structurally related purity,peak purity of thiacloprid was evaluated by five points of UV spectra.From the qNMR of hydrogen,its mass fraction value was 995.7 mg/g which was traced to the internal standard(methyl sulfone).Finally,the certified value was assigned to 997 mg/g,together with an expanded uncertainty of 3 mg/g(k=2).This high-purity thiacloprid can be applied to routine monitoring in agriculture and the food fields.
出处 《Journal of Analysis and Testing》 EI CSCD 2024年第1期1-8,共8页 分析检测(英文)
基金 financially supported by the State Administration for Market Regulation(QNBJ202306) the National Natural Science Foundation of China(22004004).