摘要 近年来,为了响应国家在医疗设备上自主创新的大力号召,涌现了一大批国产自主研发的医疗设备。然而,医疗设备的电磁兼容(EMC)设计、测试环境变得越来越复杂,医疗设备在研发过程中不可避免地产生了一系列问题。在我国医疗设备生产后要进行认证批准,才能允许销售使用。在进行认证时,根据标准YY 9706.102—2021的要求需要进行抗扰度项目测试,而静电放电(ESD)测试是其中一项重要的测试项目,其会对医疗设备的正常使用产生最直接的影响。故针对医疗设备在ESD测试时出现的故障进行了理论分析和相应的整改,整改后设备顺利通过了ESD测试,从而验证了所提供对策的可行性和实用性。 In recent years,in response to the country’s strong call for independent innovation in medical equipment,a large number of domestic independently developed medical equipment have emerged.However,the electromagnetic compatibility(EMC)design and testing environment of medical devices have become increasingly complex,and a series of problems inevitably arise during the development process of medical devices.After the production of medical equipment in China,certification and approval are required before it can be allowed for sale and use.When conducting certification,according to the requirement of YY 9706.102—2021,immunity testing is required,and electrostatic discharge(ESD)testing is one of the important testing items,it will have the most direct impact on the normal use of medical equipment.Therefore,the theoretical analysis and corresponding rectification are carried out for the faults of medical equipment in ESD test.After rectification,the equipment successfully passed ESD testing,thereby verifying the feasibility and practicality of the provided countermeasures.
作者 钱天元 杜季飞 QIAN Tianyuan;DU Jifei(Jiangsu Electronic Information Product Quality Supervision&Inspection Institute(Jiangsu Information Security Evaluation Center),Wuxi 214072,China)
出处 《电子质量》 2024年第4期86-90,共5页 Electronics Quality
关键词 医疗设备 YY 9706.102—2021 静电放电测试 整改 medical equipment YY 9706.102—2021 ESD testing rectification
  • 相关文献

参考文献3

二级参考文献27

  • 1林瑜,陈家能,陈贻焕,张芳.高速PCB板的电磁兼容设计[J].现代电子技术,2005,28(15):110-112. 被引量:10
  • 2电子式电能表静电放电的抗扰度试验[J].贵州电力技术,2005,8(10):42-43. 被引量:1
  • 3张枫,杨树彬,杨安民,吴瑞德.一种低冲击分离螺栓的设计[J].火工品,2006(3):14-17. 被引量:10
  • 4蔡瑞娇.火工品设计原理[M].北京:北京理工大学出版社,1997.
  • 5ESD STMS. 1. ESD Association Standard Test Method for Electrostatic Discharge Sensitivity Testing Human Body Model Component Level [S]. ESD Association, 2001.
  • 6JESD22-A114-B. JEDEC Standard for Electrostatic Discharge (ESD) Sensitivity Testing Human Body Modal (HBM) [S]. JEDEC Solid State Technology Association, 2001.
  • 7AEC-Q100-002-REV-F. AEC Standard for Human Body Model Electrostatic Discharge Test [S]. Automotive Electronics Council Component Technical Committee, 2003.
  • 8JOACHIM C. Relevance of Contact Reliability in HBMESD Test Equipment[J].Microelectronics Reliability , 2001,41 (9-10):1397-1401.
  • 9VERHAEGE K. Justifications for Reducing HBM and MM ESD Qualification Test Time[J].Microelectronics Reliability, 1996,36 : 1715-1718.
  • 10HUANG C Y,CHEN W F,CHUAN S Y,et al. Design Optimization of ESD Protection and Latchup Prevention for A Serial I/O IC[J].Microelectronics Reliability , 2004,44, : 213-221.

共引文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部