WYD2000型野战手术灯的维修及改进方案

Maintenance and improvement program for WYD2000 field surgical lamp
下载PDF
导出
摘要 介绍了WYD2000型野战手术灯的安装、撤收步骤,分析了WYD2000型野战手术灯在拆装过程中遇到的插头管折断故障及产生的原因,设计制作了专用维修工具,并优化改进了插头管制作材料和固定方式,解决了WYD2000型野战手术灯插头管易折断、难维修的问题,为WYD2000型野战手术灯的使用维护和改进提供了参考。 The steps for installation and withdrawal of WYD2000 field surgical lamp were introduced.The failures and causes of broken cross-arm connector of WYD2000 field surgical lamp were analyzed.The problems of WYD2000 field surgical lamp in vulnerability to breaking and difficulty in maintenance were solved by designing and manufacturing a special maintenance tool and optimizing the materials and fixing mode of cross-arm connection.References were provided for main-tenance and improvement of WYD2000 field surgical lamp.
作者 朱明明 曹冀苏 张泽瑞 郝江辉 王睿哲 周峰 荀鲁川 ZHU Ming-ming;CAO Ji-su;ZHANG Ze-rui;HAO Jiang-hui;WANG Rui-zhe;ZHOU Feng;XUN Lu-chuan(Drug and Instrument Supervision and Inspection Station of Xining Joint Logistics Support Center,Lanzhou 730050,China;The Third Procurement Service Station of Joint Logistics Support Force,Chengdu 610051,China)
出处 《医疗卫生装备》 CAS 2024年第4期116-118,共3页 Chinese Medical Equipment Journal
关键词 WYD2000型野战手术灯 故障维修 改进方案 WYD2000 field surgical lamp troubleshooting improvement program