期刊文献+

室上性心动过速机制的智能分类模型:基于十二导联穿戴式心电设备

An intelligent model for classifying supraventricular tachycardia mechanisms based on 12-lead wearable electrocardiogram devices
下载PDF
导出
摘要 目的基于十二导联穿戴式心电设备,探索室上性心动过速(SVT)机制鉴别的智能分类模型。方法选取356份SVT的穿戴式心电图,通过五折交叉验证的方式随机分为训练集、验证集建立智能分类模型,选取2021年10月~2023年3月诊断为SVT并行电生理检查及射频消融术的患者共101例作为测试集。对比心动过速诱发前后的心电图参数改变,基于多尺度深度神经网络,并加入窦性心律对比图增强训练,建立SVT机制分类的智能分类模型并验证诊断效能。进一步提取II,III,V1三导联心电信号建立分类模型,并对比其与十二导联智能分类模型的效能。结果101例测试集中68例为房室结折返性心动过速,33例为房室折返性心动过速。预训练模型在验证集中识别房室结折返性心动过速的最高精确率-召回率曲线下面积达到0.9492,F1评分为0.8195。最终II导联,III导联,V1导联,三导联与十二导联智能分类模型于测试集中的总F1评分分别为0.5597,0.6061,0.3419,0.6003与0.6136。对比十二导联,III导联的净重新分类指数与综合判别改善指数分别为-0.029(P=0.714)与-0.005(P=0.817)。结论基于多尺度深度神经网络,初步建立了穿戴式心电图对SVT机制分类的智能分类模型,并具有一定的准确性。 Objective To develop an intelligent model for differential diagnosis of atrioventricular nodal re-entrant tachycardia(AVNRT)and atrioventricular re-entrant tachycardia(AVRT)using 12-lead wearable electrocardiogram devices.Methods A total of 356 samples of 12-lead supraventricular tachycardia(SVT)electrocardiograms recorded by wearable devices were randomly divided into training and validation sets using 5-fold cross validation to establish the intelligent classification model,and 101 patients with the diagnosis of SVT undergoing electrophysiological studies and radiofrequency ablation from October,2021 to March,2023 were selected as the testing set.The changes in electrocardiogram parameters before and during induced tachycardia were compared.Based on multiscale deep neural network,an intelligent diagnosis model for classifying SVT mechanisms was constructed and validated.The 3-lead electrocardiogram signals from II,III,and V1 were extracted to build new classification models,whose diagnostic efficacy was compared with that of the 12-lead model.Results Of the 101 patients with SVT in the testing set,68 were diagnosed with AVNRT and 33 were diagnosed with AVRT by electrophysiological study.The pre-trained model achieved a high area under the precision-recall curve(0.9492)and F1 score(0.8195)for identifying AVNRT in the validation set.The total F1 scores of the lead II,III,V1,3-lead and 12-lead intelligent diagnostic models in the testing set were 0.5597,0.6061,0.3419,0.6003 and 0.6136,respectively.Compared with the 12-lead classification model,the lead-III model had a net reclassification index improvement of-0.029(P=0.878)and an integrated discrimination index improvement of-0.005(P=0.965).Conclusion The intelligent diagnostic model based on multiscale deep neural network using wearable electrocardiogram devices has an acceptable accuracy for classifying SVT mechanisms.
作者 王泓森 米利杰 张越 葛兰 赖杰伟 陈韬 李健 时向民 修建成 唐闵 阳维 郭军 WANG Hongsen;MI Lijie;ZHANG Yue;GE Lan;LAI Jiewei;CHEN Tao;LI Jian;SHI Xiangmin;XIU Jiancheng;TANG Min;YANG Wei;GUO Jun(Department of Cardiovascular Medicine,Sixth Medical Center of Chinese PLA General Hospital,Beijing 100048,China;Arrhythmia Center of Fuwai Hospital,Chinese Academy of Medical Sciences,Beijing 100037,China;School of Biomedical Engineering,Southern Medical University,Guangzhou 510515,China;Department of Cardiovascular,Nanfang Hospital,Southern Medical University,Guangzhou 510515,China)
出处 《南方医科大学学报》 CAS CSCD 北大核心 2024年第5期851-858,共8页 Journal of Southern Medical University
基金 国家重点研发计划(2018YFC2001205) 军委后勤保健专项(22BJZ41)。
关键词 穿戴式心电图 室上性心动过速 十二导联心电图 多尺度深度神经网络 房室结折返性心动过速 房室折返性心动过速 wearable electrocardiogram supraventricular tachycardia 12-lead electrocardiogram multiscale deep neural network atrioventricular nodal re-entrant tachycardia atrioventricular re-entrant tachycardia
  • 相关文献

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部